18-2
IM AQ1200-01EN
18.2 Error Messages
Error Messages
Messages may appear on the screen while you are using the AQ1200. This section explains the
meanings of the error messages and how to respond to them. You can display the error messages in
English or Japanese (see section 17.1). If servicing is necessary to solve the problem indicated by a
message, contact your nearest YOKOGAWA dealer.
Other communication-related error messages may also appear. These error messages are described
in the
Communication Interface User’s Manual
, IM AQ1200-17EN.
Error in Execution
Code
Message
20
Network Option is not installed.
24
One or more conditions in this file are not supported by this product.
25
The Real Time measurement is executed. Thus, the setting mode has been changed from Multi-wavelength Mode to
Advanced Mode.
27
The end point is different between acquired data and reference data. Please check the connected fiber cable.
28
The measurement may not be completed within the specified duration.
29
In remote control mode, all keys are locked except F1 key. Please hit F1 key to exit the remote control mode.
33
File operation is cancelled.
35
The optical plug may not be connected securely.
36
Either of the analysis information below could not be retrieved. (Cursor info, Free marker info., Event info.)
37
The end point distance does not match with the master end point.
40
No more averaging is allowed. (already MAX averaging condition) Set "Average Continue"OFF for new
measurement.
41
Now updating firmware. Please wait. It takes approx. 20 seconds. Do not turn off power during update. Restart
automatically after successful updating.
64
Failed to adjust LS power. Please check whether the patch fiber cable is properly connected between LS port and
OPM port.
65
Project Information doesn’t match between Master and Slave.
66
The project sent from master includes an unsuitable wavelength.
500
The measurement condition is failed. Please reset ****** in SETTING menu.
501
Not executable during measurement. Please stop the measurement and execute again.
502
The measurement cannot be started during printing. Please either wait until the end of the printing or interrupt the
printing.
503
The measurement cannot be started during the file operation. Please either interrupt the file operation or wait until
the end of the file operation.
504
The measurement could not be completed within the specified averaging duration. Please modify the averaging
duration.
505
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
506
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
507
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
508
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
509
Hardware failed, and needs to be repaired. Please contact Yokogawa’s representatives.
512
ZERO SET ERROR
513
ZERO SET ERROR
514
Exceeding limit. It may cause damage of the instrument. Please disconnect the plug.
515
ZERO SET ERROR
516
Fiber In Use Alarm Error 1
517
Fiber In Use Alarm Error 2
601
Cannot save onto the media. This media is for read-only.
602, 603
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
604
Storage media has been disconnected while the media is being accessed.
605
Same file name or folder name exists.
606, 607
Can not recognize file system. - Internal memory:Contact Yokogawa’s representatives. - USB memory:Try the other
media or format again with FAT.
608
Invalid file name or folder name