<6.  OUTLINE>

6-5

IM 01E25D11-01EN

n

 OPTIONAL SPECIFICATIONS 

 : Available   – : Not available

Item

Specifications

Applicable Model

Code

General

Submersible

AXW***G AXW***W

Grounding Rings (*1) JIS SUS304 (AISI 304 SS Equivalent), a set of two rings

GR1

Grounding Rings (*2) SUSF304 (AISI 304 SS Equivalent), a set of two rings with gaskets “B”.

GR2

For District Heating 

and Cooling or  

Condensation-proof

Urethane resin potting is applied in the terminal box of a remote flowtube. Select JIS 

G1/2 for the electrical connections. 30-meter dedicated signal and excitation cables are 

pre-wired and waterproof glands with union joints are attached at factory.

DHC

User-specified Signal 

and Excitation Cable 

Length

Available for the submersible type and a model with optional code DHC. The cable 

length is limited up to 200 meters combined with an AXFA11 converter. Following “L,” 

specify the cable length in three digits as a multiple of 1 meter (e. g., 001, 002, or 005) 

for a length up to 5 meters, or as a multiple of 5 meters (i.e., 005, 010, 015, or the like) 

for a length of 5 meters or more. If this optional code is not selected, a 30-meter length 

cable is attached.

L***

Mass Unit Setting

The flow rate span, transmission pulse weight, and totalizer display pulse weight can be 

set in terms of mass unit.

Specify the density of the process fluid when ordering in addition to the mass flow rate 

span, transmission pulse weight (for mass unit), and totalizer display pulse weight (for 

mass unit).

When ordering a remote flowtube, parameters for 'Mass Unit Setting' will be set in the 

corresponding converter before shipment.

1. Density

a. Available density Numerics:

Specify the numeric within the value of 500 to 2000 kg/m

3

, 4.2 to 16.7 lb/gal, or 31.2 

to 124.8 lb/cf. And it can be up to five digits, to a maximum of 32000 ignoring the 

decimal point. A fraction is limited to the fourth decimal place.

b. Available density units: kg/m

3

, lb/gal, lb/cf

Example: A water density is about 1000 kg/m

3

. In this case specify “1000kg/m

3

”.

However a density is changed by temperature. Specify the actual density.

(The 1000 kg/m

3

 is equivalent to 8.345lb/gal and 62.43lb/cf.)

2.  The mass flow rate span, transmission pulse weight, and totalizer display pulse 

weight

a. Available density Numerics:

Specify the numeric within the value of 0.0001 to 32000. And it can be up to five 

digits, to a maximum of 32000 ignoring the decimal point. A fraction is limited to the 

fourth decimal place.

b. Mass Units  

 Available mass units: t, kg, g, klb, lb  

Available time units: /d, /h, /min, /s

Note1:  In case of specifying the mass flow span, calculate the volumetric flow span 

by the setting density, and specify the available value in the mass flow span.

Note2:  In case of transmission pulse weight and totalizer display pulse weight, 

specify the mass unit which was specified as the flow unit.

MU

G3/4 Female

Waterproof Glands

Waterproof glands for G3/4 conduits or flexible tubes are attached to the electrical con-

nections. Available only for JIS G1/2 electric connections.

EW

Waterproof Glands

Waterproof glands are attached to the electrical connections. Available only for JIS G1/2 

electric connections.

EG

Waterproof Glands

with Union Joints

Waterproof glands with union joints are attached to the electrical connections. Available 

only for JIS G1/2 electric connections.

EU

Stainless Steel Tag 

Plate

A pendant tag plate of stainless steel (AISI 304 SS equivalent) is provided in addition to 

the standard nameplate with the tag number inscribed on it.

Dimension (Height x Width): Approx. 12.5 (4.92) x 40 (15.7) mm (inch)

SCT

Direction Change of 

Electrical  

Connection (*3)

+90 degrees rotated terminal box to change the direction of the electrical connection.

RA

+180 degrees rotated terminal box to change the direction of the electrical connection.

RB

–90 degrees rotated terminal box to change the direction of the electrical connection.

RC

Material Certificate

Material certificates are provided for pipe, electrodes, flanges, and also optional ground-

ing rings when specified.

M01

*1:  This is not available for models with process connection JIS F12 (JIS 75M) of sizes  1100 mm (44 in.) to 1800 mm (72 in.). Refer to the section 

on “Mounting Procedures” on necessary gaskets.

*2:  This is available only for models with process connection JIS F12 (JIS 75M) of sizes 1100 mm (44 in.) to 1800 mm (72 in.). Refer to the section 

on “Mounting Procedures” on necessary gaskets.

*3:  Refer to the image below.

Standard

+90 degree rotation

+180 degree rotation

–90 degree rotation

Optional Code RA

Optional Code RB

Optional Code RC

Electrical

Connection

Electrical

Connection

Summary of Contents for Admag AXW

Page 1: ...User s Manual AXW Magnetic Flowmeter Remote Flowtube IM 01E25D11 01EN IM 01E25D11 01EN 2nd Edition...

Page 2: ...ns 3 1 3 2 Handling Precautions 3 3 3 2 1 General Precautions 3 3 3 2 2 Flowmeter Piping 3 4 3 3 Mounting Procedures 3 4 4 WIRING 4 1 4 1 Wiring the Remote Flowtube 4 1 4 1 1 Wiring Precautions 4 1 4...

Page 3: ...ii IM 01E25D11 01EN n OPTIONAL SPECIFICATIONS 6 5 n OPTIONAL SPECIFICATIONS continued 6 6 n DIMENSIONAL DRAWINGS 6 7 n SIZING DATA 6 19 Revision Information i...

Page 4: ...t changes that are not considered to affect function or performance If the customer or any third party is harmed by the use of this product Yokogawa assumes no responsibility for any such harm owing t...

Page 5: ...2 Wiring WARNING The wiring of the magnetic flowmeter must be performed by expert engineer or skilled personnel No operator shall be permitted to perform procedures relating to wiring When connecting...

Page 6: ...ould appreciate the inclusion of drawings or additional information The results of our examination will determine whether the meter will be repaired free of charge or on an at cost basis The guarantee...

Page 7: ...Blank Page...

Page 8: ...A place where it will not be exposed to rain or water A place subject to minimal vibrations or shocks Temperature and humidity levels should be as follows Temperature 10 to 70 C Humidity 5 to 80 RH n...

Page 9: ...Blank Page...

Page 10: ...ng two or more magnetic flowmeters on a single pipe provide a run of at least 10D between them F0301 ai D Flowtube Size Gate valve fully open 5D or more 2D or more Reducer pipe Expander pipe 0 is allo...

Page 11: ...r to the pipeline set a expansive pipe in the downstream side 9 Mounting Positions Pipes must be fully filled with liquids IMPORTANT It is essential that pipes remain fully filled at all times otherwi...

Page 12: ...ed to the installing location transport it to the site in its original packing CAUTION When lifting the flowtube use the lifting rings eye bolts or eye plates as in Figure 3 2 1 Never lift it using a...

Page 13: ...substances passing through the flowtube during measurement 3 3 Mounting Procedures CAUTION When attaching optional grounding rings to the flowtube with lining material natural hard rubber or fluoroca...

Page 14: ...of gasket should be a kind of soft rubber or its equivalent in hardness 1 Standard no grounding rings Gaskets A supplied by customer Customer pipe Customer pipe A A F0310 ai Figure 3 3 2 Installation...

Page 15: ...of the flowtube s flange and fix the wire by the nut This procedure item 1 to 4 must be done for the both sides of the flowtube 5 Install the flowtube into the customer s pipe with the gaskets A F0313...

Page 16: ...Procedure with Grounding Rings GR1 for sizes 1100 mm 44 in and above Mounting Procedure with optional Grounding Rings GR2 The grounding rings are installed and locked onto the flowtube with gaskets B...

Page 17: ...ess connection Size mm inch ASME B16 5 500 600 B16 47 700 1000 Class 150 EN1092 1 PN10 JIS B2220 10K EN 1092 1 PN16 AWWA C207 Class D AS2129 Table D E AS4087 PN16 JIS F12 JIS 75M 500 20 71 to 118 60 t...

Page 18: ...mm inch ASME B16 5 500 600 B16 47 700 1000 Class 150 EN1092 1 PN10 JIS B2220 10K EN 1092 1 PN16 AWWA C207 Class D AS2129 Table D E AS4087 PN16 JIS F12 JIS 75M 500 20 626 to 1044 527 to 876 652 to 108...

Page 19: ...Blank Page...

Page 20: ...gth All the cable ends must be provided with round crimp on terminals and be securely wired The signal cables must be routed in separate steel conduit tubes 16 JIS C 8305 or flexible conduit tubes 15...

Page 21: ...rter side On the flowtube side Figure 4 1 2 Treatment of Dedicated Signal Cables CAUTION As crimp terminals A B SA SB and C have their own electrical potentials securely insulate them so as not to com...

Page 22: ...ing at this terminal box is not necessary and do not open it IMPORTANT The wiring port is sealed with a cap not water proof Do not remove the cap from the wiring port until wiring a cable If waterproo...

Page 23: ...regularly Drain valve F0406 ai Figure 4 1 6 Conduit Wiring 4 1 4 Wiring Connections WARNING Before wiring be sure that the power supply for AXFA11 converter has been turned off to prevent an electric...

Page 24: ...grounding can have an adverse affect on the flow measurement Ensure that the instrument is properly grounded The electromotive force of the magnetic flowmeter is minute and it is easy to be affected...

Page 25: ...Blank Page...

Page 26: ...ot be changed direction of electrical connection after delivery Submersible Type Optional code DHC for district heating and cooling or condensation proof 1 The following tools are required to change t...

Page 27: ...an abnormal condition Consult Yokogawa s sales or service offices 2 Insulation Resistance Check Check the insulation resistances in the terminal box in accordance with the tables below If any of them...

Page 28: ...flowtube is filled with fluid Is the flowtube full of fluid Remove the tube and inspect the inside of the pipe Is foreign matter adhering to the walls Does the arrow on the tube match the flow directi...

Page 29: ...imits Whenever possible reduce the flow velocity to being in the vicinity of the lower conductivity limit Remove dirt and scale of and near electrodes Use a soft brush or cloth to perform scale remova...

Page 30: ...ed by turbulence in the flow Relocate the valve to the downstream side If the valve must be located on the upstream side ensure that the minimum distance to the flowmeter is 10 times the flow tube dia...

Page 31: ...Blank Page...

Page 32: ...een Submersible type Pipe Housing Flange Terminal Box Non tar epoxy coating black Wiring Electrical connection ANSI 1 2 NPT female ISO M20 x 1 5 female JIS G1 2 PF1 2 female Direction of Electrical Co...

Page 33: ...rature 10 to 60 C 14 to 140 F Note Minimum temperature should also be limited accord ing to minimum fluid temperature of flow tube s speci fication Refer to description of Fluid Temperature and Pressu...

Page 34: ...p to 0 6 MPa CE1 EN1092 1 PN10 Flange Carbon Steel fluid pressure up to 1 MPa 500 mm 20 in to 1000 mm 40 in for lining code F H D U 1200 mm 48 in to 1800 mm 72 in for lining code H CE2 EN1092 1 PN16 F...

Page 35: ...G1 2 female 2 ANSI 1 2 NPT female Not available for Submersible type 4 ISO M20 x 1 5 female Not available for Submersible type Indicator N None Calibration B Standard Optional Specifications Optional...

Page 36: ...zer display pulse weight a Available density Numerics Specify the numeric within the value of 0 0001 to 32000 And it can be up to five digits to a maximum of 32000 ignoring the decimal point A fractio...

Page 37: ...ard List are issued L3 Level 4 The Declaration and the Yokogawa Measuring Instruments Control System are issued L4 Five point Calibra tion in User specified Span A flow test near 0 25 50 75 and 100 of...

Page 38: ...749 3 29 50 863 6 34 00 977 9 38 50 1086 42 75 1200 47 25 Inner Diameter d Lining H F D 477 18 78 578 22 76 675 26 57 777 30 59 878 34 57 976 38 43 Lining U 469 18 46 570 22 44 667 26 26 769 30 28 870...

Page 39: ...ble 3 EN PN10 Flange Type Unit mm approx in Process Connection Code CE1 Flange Type EN1092 1 PN10 Size Code 500 600 700 800 900 10L Size 500 20 600 24 700 28 800 32 900 36 1000 40 Lining Code H F U D...

Page 40: ...ion code DHC is selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 5 JIS 10K Flange Type Unit mm approx in Process Connection Code CJ1 Fl...

Page 41: ...s selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 7 AS2129 Table E Flange Type Unit mm approx in Process Connection Code CS2 Flange Ty...

Page 42: ...submersible type or option code DHC is selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 9 JIS F12 JIS 75M Flange Type Unit mm approx in...

Page 43: ...24 700 28 800 32 900 36 1000 40 Lining Code H F U D Ring Outer Diameter D 590 23 23 690 27 17 799 31 46 906 35 67 1006 39 61 1116 43 94 Ring Inner Diameter d 485 19 09 586 23 07 683 26 89 787 30 98 8...

Page 44: ...ion Code CB1 Flange Type AWWA C207 Class D Size Code 12L Size 1200 48 Lining Code H Lay Length L 1440 56 69 Flange Outer Diameter D 1511 3 59 50 Flange Bolt Circle Diameter C1 1422 4 56 00 Ring Bolt C...

Page 45: ...rproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 3 EN PN10 Flange Type fluid pressure up to 0 6 MPa Unit mm approx in Process Connection Code CEK Fl...

Page 46: ...type or option code DHC is selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 5 AS2129 Table D AS2129 Table E AS4087 PN16 Flange Type Uni...

Page 47: ...Diameter d Lining U 1071 42 17 1172 46 14 1322 52 05 1469 57 83 1565 61 61 1759 69 25 Bolt Hole Pitch Half Angle 7 5 6 4 6 4 5 6 5 4 1 Bolt Hole Diameter h 33 1 30 33 1 30 39 1 54 39 1 54 39 1 54 39...

Page 48: ...516 59 69 1716 67 56 1925 75 79 1335 52 56 1540 60 63 1766 69 53 1966 77 40 Ring Inner Diameter d 1195 47 05 1201 47 28 1404 55 28 1592 62 68 1790 70 47 1195 47 05 1396 54 96 1588 62 52 1782 70 16 Bol...

Page 49: ...Size Code 11L 12L 13L 15L 16L 18L Size 1100 44 1200 48 1350 54 1500 60 1600 64 1800 72 Lining Code U Thickness 25 0 98 Approx Weight Unit kg lb 59 3 131 65 7 145 75 8 167 88 3 195 92 4 204 114 2 252 O...

Page 50: ...95 0 08 2 3 0 09 2 7 0 11 3 15 0 12 3 6 0 14 4 05 0 16 4 45 0 18 4 85 0 19 500 19 69 630 24 80 800 31 50 1000 39 37 1250 49 21 1600 62 99 2000 78 74 2500 98 43 630 24 80 800 31 50 1000 39 37 1250 49...

Page 51: ...Blank Page...

Page 52: ...le AXW Magnetic Flowmeter Remote Flowtube Manual No IM 01E25D11 01EN Edition Data Page Revised Item 1st Mar 2012 New publication 2nd Sep 2013 Addition of sizes from 1100 mm 44 in to 1800 mm 72 in Addi...

Reviews: