IM 91051-01EN (P2)

2. Components

Name

Description

1. Pressure/Vacuum 

knob

Turn the knob to switch between 

 

Pressure

Vacuum

, and 

Vent

 open.

2. Handle

Operate (push) the handle to generate 

pressure.

3. Vernier adjustment

Used to “fine tune” the pressure to the 

target value.

4. Pressure output port 

1/8" NPT

Port for outputting pressure.

Connect a standard pressure meter, 

calibration target (connector), 

 

and so on to this port.

VENT

PRESSUER

VACUUM

Vent

(vent open) Pressure

Vacuum

3. Vernier adjustment

1. Pressure/Vacuum

    knob

4. Pressure output 

    port

2. Handle

3.  Operating Instructions

3.1  Handling Precautions

Be sure to check the operating environment (safety).
Before use, be sure to check that connectors are secure, 

 

that there are no intrusion of foreign substances, 

 

and that there are no damages. 
In addition, use the same thread specifications for all connectors.
Perform preliminary pressure tests on tubing and other parts of 

 

the system. 
If you are applying pressure using a hand pump, 

 

be sure to monitor the pressure with a standard pressure meter 
(monitor).

3.2  Connecting Connectors

CAUTION

 

• Use connectors appropriate for the thread specifications of 

 

the device under test (DUT) and tubing. 

  Otherwise, leakage and damage to threads may result.
• Apply appropriate seal tape to the connecting surface of 

 

the connector.

1/8” NPT

Standard pressure 

meter

Connector

The device under test

(DUT)

<Connection example>

■ Using Two Wrenches

To prevent leakage, you must tighten the connector or 

 

quick adapter firmly.
Using a wrench only on the connector side may break 

 

the pressure pump.
Use two wrenches to prevent force from being applied to 

 

the pressure pump.
(The same applies when you connect a connector to 

 

the standard pressure meter.)

3.3  Generating Pressure

WARNING

 

• Remove tubing and wiring from the the device under test (DUT) 

before carrying out the procedure.

• Only connect to this pressure pump. 
  Do not connect to another high pressure source.
• Be sure to observe all safety precautions to avoid injury and 

 

damage to the DUT.

CAUTION

 

Use connectors and hoses appropriate for the source pressure 

 

value.

[ Procedure ]

(1)  Connect the standard pressure meter and the DUT.
(2)  Use 

the Pressure/Vacuum knob

 to select pressure or 

 

vacuum.

(3)   Push 

the handle

 to apply pressure near the target pressure 

 

(operate the handle while checking the standard pressure 
meter).

(5)   Turn 

the vernier adjustment

 to “fine tune” the pressure to 

the target value. 

 

  

Increase:   Turn clockwise 

 

  

Decrease:  Turn counterclockwise

Note 

The vernier adjustment can be turned about 20 times 

 

counterclockwise (decreasing pressure) from the maximum 

 

position. If the adjuster feels heavy to turn, stop turning it. 
Turning it by force can cause damage.

3.4  Releasing Pressure

WARNING

 

To prevent accidents, safely return the pressure to zero 

 

(release the pressure) before removing the connector and hose.

After generating pressure, carefully turn 

the pressure/vacuum 

knob

 to the vent (vent open) position to release the pressure.

Note 

After use, store the pressure pump in the hand pump case or 

 

in a safe place.

4. Troubleshooting

If you cannot increase the pressure with the hand pump 

 

(if the pressure decreases), check the following.

• Check the pressure/vacuum knob position.

• Check the thread specifications of the connection port.

 

Check that the connection is secure (use seal tape or 

 

other measures).

CAUTION

 

If you cannot determine the cause, continuing with the procedure 
is dangerous. Stop immediately.
If the product does not work properly, contact your nearest 

 

YOKOGAWA dealer.