Ethernet Interface Specifications
Connector type:
RJ-45
Number of ports:
1
Electrical and mechanical specifications:
Conforms to IEEE 802.3.
Transmission system:
100BASE-TX/10BASE-T
Maximum transmission rate:
100 Mbps
Protocol:
TCP/IP
Command control server function:
A command mnemonic stream parsing server at
port 7655 that allows up to five clients to connect.
The terminator (CR, LF, or CR+LF) can be
specified.
FTP server function:
An anonymous FTP server at port 21 that allows up
to five clients to connect.
HTTP server function:
A function that enables various information to be
displayed and performs remote control of the
GS610 through port 80 using a Web browser.
Ethernet Interface Functions
Reception Function
• You can specify the same settings as those
specified by front panel key operations.
• Receives output requests for measured and
computed data, setup data of the panel, and error
codes.
Transmission Function
• Outputs measured and computed data.
• Outputs panel setup data and the status byte.
• Outputs error codes that have occurred.
Switching between Remote and Local
Modes
When Switching from Local to Remote Mode
Remote mode is activated when the :SYSTem:REMote
command is received from a controller while local
mode is active.
• The REMOTE indicator (see section 1.2) is turned
ON.
• All keys except the
LOCAL
key are disabled.
• Settings entered in local mode are retained even
when the GS610 switches to remote mode.
When Switching from Remote to Local Mode
Pressing
LOCAL
in remote mode puts the instrument
in local mode. The GS610 switches to local mode
when it receives a :SYSTem:LOCal command from the
PC.
• The REMOTE indicator is turned OFF.
• Key operations are enabled.
• Settings entered in remote mode are retained even
when the GS610 switches to local mode.