3. Joining the Instruments with Mounting Hardware

1. With the eigh labels (1), seal the eigh openings on the bottom plate to fix the feet and the side panels to fix the

handles.

2. Using two flat-head screws (M5, 14 mm long) (3), screw the left side bracket (2) to the instrument to be located

on the left side of the rack.

Reference torque value: 2.8 to 2.9 N·m

3. Using two other flat-head screws (M5, 14 mm long) (3), screw the right side bracket (4) to the right side of the

instrument to be located on the right side of the rack.

Reference torque value: 2.8 to 2.9 N·m

4. Screw the upper bracket (5) to the left and right side brackets with two flat-head screws (6) (M3, 8 mm long) .

Reference torque value: 0.6 to 0.7 N·m

1

1

5

1

2

1

3

1

4

3

6

6

1

1

4. Mounting the Joined Instruments to the Rack

Dimensions after mounting to the rack are illustrated below.

Note: 1. Be sure to place support beneath the rack where the support will not cover the vent openings

on the bottom of the instruments.

2. Provide a clearance of at least 100 mm around the instrument for ventilation.

6

17.1

Mounting surface

460

UNIT : mm

24.5

100

149

–0.3

0

480

±

1

IM 751535-J3-11E    (2/2)