IM 751533-J4-01E 2/2
3.
Joining the Right and Left Instruments with Mounting Hardware
1. With the six labels (1), seal the six openings on the bottom plate to fix the feet and the left side panel to
fix the handle.
2. Using three flat-head screws (M5, 14 mm long) (3), screw the left side bracket (2) to the left side of the
instrument.
Reference torque value: 2.8 to 2.9 N•m
3. Using three binding-head screws (M5, 14 mm long) (5), screw the right side bracket (4) to the right side
of the instrument.
Reference torque value: 2.8 to 2.9 N•m
4. Screw the upper bracket (6) to the left side bracket (2) with a flat-head screw (M3, 8 mm long) (8) and to
the right side bracket (4) with two binding-head screws (M3, 8 mm long) (7) in the appropriate positions.
Reference torque value: 2.8 to 2.9 N•m
SETUP
ALL
ESC
RESET
ELEME
NT
1
2
3
ELEMEN
T
RANGE
VOLTA
GE
CURRE
NT
AUTO
AUTO
DISPL
AY
NUMERIC
CURSO
R
START
/
STOP
RESET
SINGLE
NULL
KEY LO
CK
FILE
IMAGE
MENU
STORE
STORE
SET
POW
ER
SET
CAL
PAGE
PAGE
WAVE
OTHERS
FORM
HOLD
LOCAL
MISC
SHIFT
ITEM
INPUT IN
FO
INTEGRA
TOR
5
3
4
2
1
1
1
1
1
7
6
7
8
4. Mounting the Joined Instruments to the Rack
Dimensions after mounting to the rack are illustrated below.
Note:
1. Be sure to support the instrument from the bottom. When doing so, make sure not to cover the
ventilation holes on the bottom of the instrument.
2. Provide a clearance of at least 100 mm around the instrument to prevent the temperature rise
inside the instrument.
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3 mm.
Rack mount attachment plane
6
460
480
±1
50
50
50
24.5
199
-0.3
28.6
(0.24
)
(18.11)
(18.90
±0.04
)
(0.96
)
( 1.97
)
( 1.97
)
( 1.97
)
(7.83
-0.01
)
(1.13)
0
0