YOKOGAWA 702907 User Manual Download Page 2

French

 ATTENTION

N’appliquez pas une force excessive sur la vis de réglage de phase. Cela pourrait 

endommager le condensateur variable interne.

1. 

Connect the BNC connector of the probe to the input of the oscilloscope, and connect the probe 

input to the oscilloscope signal output terminal for probe compensation adjustment (CAL or 

COMP terminal).

2. 

Operate the vertical and time scales on the oscilloscope, and turn the phase adjustment screw 

with the included adjustment screwdriver to adjust the observed waveform to the correct 

waveform (square wave). Adjust the probe capacitance to match the input capacitance of the 

oscilloscope with the variable capacitor inside the probe.

 

Correct Waveform Inadequate Compensation Over Compensation

Note

Accurate measurements may not be possible near objects with strong electromagnetic fields such 

as transformers, large current circuits, or wireless equipment.

4. Specifications

Item

Specifications

Probe length

2500 mm ±50 mm (including the pincher tip)

Weight

70 g (probe only)

Connector type

BNC (with the probe ID pin)

Attenuation ratio

10:1 ±2 % (+5 °C to +40 °C)

1

10:1 ±3 % (−40 °C to +5 °C, +40 °C to +85 °C)

1

Frequency band

DC to 200 MHz

 

(−3 dB)

2

Rise time

1.75 ns (typical)

2,3

Propagation delay time

12.0 ns (typical)

3

Maximum input voltage

±1000 V (DC + ACpeak) (EN 61010-031)

4

Input resistance

10 MΩ ±2%

1

Input capacitance

18.0 pF (typical)

3

Matching input capacitance

12 pF to 22 pF (typical)

3

Operating 

environment

Temperature/Humidity

−40 °C to 0 °C (no condensation).

 

0 °C to +40 °C, 80 %RH

5

 (no condensation). 

+40 °C to +85 °C, Relative humidity decreases from 80 

%RH at +40 °C with temperature

5

 (no condensation). 

However, up to +40 °C on the phase adjustment unit.

Altitude

3000 m or less

Storage 

environment

Temperature/Humidity

−40 °C to +85 °C, Relative humidity decreases from 80 

%RH at +40 °C with temperature

5

 (no condensation)

Altitude

4600 m or less

Compliant 

standards

Safety standard

EN 61010-031

Measurement category II

6

  1000 V (DC + ACpeak)

Pollution degree 2

7

Environmental standards

8

EU RoHS Directive compliant

1  In combination with an oscilloscope with an input impedance of 1 MΩ ±1%.

2  Varies depending on the oscilloscope combined.

3 “Typical” values are typical or average values and are not strictly guaranteed.

4  Refer to “Compliant standards (Safety standards)”. See the figure below for the input voltage 

derating by frequency.

5  See the figure below for temperature/humidity derating.

6  The product is for measurement category II (CAT II). Do not use it with measurement category 

III (CAT III), nor measurement category IV (CAT IV). When using devices or accessories with 

different measurement categories, the lower measurement category applies. See below for 

definitions of measurement categories.

Measurement Category

Definition

Measurement category “O (Other)”

Measurement category O (Other) applies to 

measurement of a circuit that is not connected directly 

to the main power source.

Measurement category II (CAT II)

CAT II applies to measurement of electrical equipment 

that is powered through a fixed installation such 

as a wall outlet wired to a distribution board and 

measurement on such wiring.

Measurement category III (CAT III)

CAT III applies to measurement at the distribution level, 

that is, building wiring, fixed installations.

Measurement category IV (CAT IV)

CAT IV applies to measurement at the primary supply 

level, that is, overhead lines, cable systems.

7  Pollution degree applies to the degree of adhesion of a solid, liquid, or gas which deteriorates 

withstand voltage or surface resistivity. Pollution Degree 2 applies to normal indoor atmospheres 

(usually with only non-conductive pollution).

8  For conformity to environmental regulations and/or standards other than EU, contact your 

nearest Yokogawa office (PIM 113-01Z2).

Input Voltage Derating by Frequency and Temperature/humidity Derating

 WARNING

As the frequency of the input signal increases, the maximum input voltage of the probe 

decreases.

French

 AVERTISSEMENT

Lorsque la fréquence du signal d’entrée augmente, la tension d’entrée maximale de la sonde 

diminue.

 

Input Voltage Derating by Frequency

10000

1000

100

10

0.01

0.1

1

10

100

Frequency [MHz]

Maximum input voltage 

[V]

20 V

200

Temperature vs. Humidity Derating

Temperature [°C]

Humidity [% RH]

90
80
70
60
50
40
30
20
10

0

90

80

70

60

50

40

30

20

10

0

100

85

11.6 %RH

Temperature Cycle Reference Values

  It has been verified that the probe can withstand at least 500 repetitions of the following temperature 

cycle in a resting state. However, these are reference values and are not strictly guaranteed.

 

[°C]

+85

-40

0

(1) (2) (3) (4)

[min]

One cycle duration

(1) 

40 

°C

 to +85 

°C : 40 min to 50 min

(2) +85 

°C                  : 

> 30 min

(3) +85 

°C to 

40 

°C : 80 min to 90 min

(4) 

40 

°C

                  : > 5 min

3. How to Use

 WARNING

•  Use this probe only with YOKOGAWA’s oscilloscopes. Even with YOKOGAWA’s 

oscilloscopes, this probe can be used only when specified as a connectable accessory. 

Also, use this probe only with standard accessories or optional accessories sold 

separately.

•  When connecting the probe to the object under measurement, be careful of electric 

shock, burns, and frostbite.

•  Do not apply a voltage exceeding the maximum input voltage to the probe. EN 61010-

031 is a safety standard that applies to the probe alone. For the actual compliant safety 

standards and operating conditions, observe the conditions for the oscilloscope. Failure 

to observe them may cause accidents, such as electric shock and damage to the 

instrument.

•  When the oscilloscope’s input coupling is AC, a DC voltage is applied to the 

oscilloscope’s input at the same electric potential as the probe’s input. Make sure not to 

exceed the oscilloscope’s maximum input voltage.

•  This probe cannot be used to measure the voltage between two points floating from the 

ground potential. Consider using a differential probe.

•  After use, turn off the power to the device under measurement, disconnect the probe from 

the device under measurement, and then disconnect the probe from the oscilloscope.

 CAUTION

•  When using optional tip accessories (such as alligator clip adapters) sold separately, 

make sure to use the 4Φ conversion adapters sold separately in between. Optional tip 

accessories are unavailabe without the 4Φ conversion adapters.

•  When connecting the probe tip directly to the device under measurement without using 

the attached pincher tip or optional accessories, use 40 V or less.

•  The product is not dust proof or water resistant. Do not use the product in areas with a lot 

of dust or where water may be spilled.

•  Avoid using or storing the product in an environment that does not meet the specifications 

such as under direct sunlight, high temperature, high humidity, or where condensation 

may form. Deformation and insulation deterioration can occur to cause the probe to no 

longer meet the specifications.

•  Avoid vibration, shock, and static electricity when handling the probe. Do not bend or pull 

the cables excessively. Doing so may damage or disconnect the probe.

•  When cleaning the probe, wipe with a piece of soft cloth to prevent damaging the probe. 

Do not immerse the probe body in liquid. Do not use abrasive cleaners or volatile solvents 

such as benzine on the probe.

French

 AVERTISSEMENT

•  Utilisez la sonde uniquement avec les oscilloscopes de YOKOGAWA. Même avec les 

oscilloscopes de YOKOGAWA, la sonde ne peut être utilisée que si spécifiée comme 

accessoire connectable. Et utilisez la sonde uniquement avec des accessoires standard 

ou des accessoires en option vendus séparément.

•  Lors de la connexion de la sonde à l’objet faisant l’objet de la mesure, faire attention au 

choc électrique, aux brûlures et aux gelures.

•  N’appliquez pas une tension supérieure à la tension d’entrée maximale sur la sonde. 

EN61010-031 est une norme de sécurité conforme qui s’applique à la sonde seule. Pour 

les normes de sécurité et les conditions d’utilisation actuelles, suivez les conditions de 

l’oscilloscope. Si cette précaution n’est pas prise, des accidents tels qu’un choc électrique 

ou un dégât matériel peuvent se produire.

•  Lorsque le couplage d’entrée de l’oscilloscope est AC, une tension DC est appliquée à 

l’entrée de l’oscilloscope au même potentiel électrique que l’entrée de la sonde. Assurez-

vous de ne pas dépasser la tension d’entrée maximale de l’oscilloscope.

•  La sonde ne peut être utilisée pour mesurer la tension entre deux points flottant du 

potentiel de terre. Pensez à utiliser une sonde différentielle.

•  Après utilisation, coupez l’alimentation de l’appareil en cours de mesure, déconnectez la 

sonde de l’appareil en cours de mesure, puis déconnectez la sonde de l’oscilloscope.

 ATTENTION

•  Lorsque vous utilisez des accessoires de pointe en option (tels que des adaptateurs 

de pince crocodile) vendus séparément, assurez-vous d’utiliser les adaptateurs de 

conversion 4Φ vendus séparément entre les deux. Les accessoires de pointe en option 

ne sont pas disponibles sans les adaptateurs de conversion 4Φ.

•  Lorsque vous connectez la pointe de la sonde directement à l’appareil sous mesure sans 

utiliser la pointe de pince attachée ou les accessoires en option, utilisez 40 V ou moins.

•  Le produit n’est pas étanche à la poussière ni à l’eau. N’utilisez pas le produit dans des 

zones très poussiéreuses ou dans lesquelles de l’eau pourrait être renversée.

•  Évitez d’utiliser ou de stocker le produit dans un environnement qui ne répond pas aux 

spécifications, par exemple sous la lumière directe du soleil, à des températures élevées, 

à une humidité élevée ou là où de la condensation peut se former. Une déformation et 

une détérioration de l’isolation peuvent survenir, faisant que la sonde ne répond plus aux 

spécifications.

•  Évitez les vibrations, les chocs et l’électricité statique lors de la manipulation du produit. 

Ne pliez pas et ne tirez pas les câbles de manière excessive. Cela pourrait endommager 

ou déconnecter la sonde.

•  Lors du nettoyage de la sonde, essuyez avec un morceau de chiffon doux pour éviter 

d’endommager la sonde. Ne plongez pas le corps de la sonde dans un liquide. N’utilisez 

pas de nettoyants abrasifs ni de solvants volatils tels que la benzine sur la sonde.

How to Connect

1. 

Connect the BNC connector of the probe to the input of the oscilloscope with the input 

impedances of 1 MΩ. Set the input impedance of the oscilloscope to 1 MΩ.

2. 

The probe ID pin is automatically detected, and the oscilloscope attenuation ratio is automatically 

set to 10:1. If it is not set automatically, set it manually.

3. 

Depending on the object under measurement or the situations, connect the attached pincher tip 

and safety ground lead to the tips of the probe, or connect the optional accessories by putting 

the optional 4Φ conversion adapters (signal end / ground) in between. See Connecting the 

Optional Accessory for details.

4. 

Adjust the phase of the probe using the screwdriver. See Phase Adjustment for details.

Connecting the Optional Accessories

  If you use the optional 4Φ conversion adapters (signal end B8099NL / ground B8099NM, 1000 

Vpeak CAT II) instead of the attached pincher tip (B8099MR) and safety ground lead (B8099NK), 

you can connect to the following optional tip accessories. We offer a variety of combinations. 

Contact your nearest YOKOGAWA dealer for details.

•  Optional pincher tip (B9852MN/B9852MM, 1000 Vrms CAT III)

•  Alligator clip adapter (758929, 1000 Vrms CAT II)

•  Fork terminal adapter (758921, 1000 Vrms CAT II)

 

Alligator clip adapter

 (red)

4Φ conversion adapter (signal end)

4Φ conversion adapter (ground)

Alligator clip adapter (black)

Connection example with accessories

Phase Adjustment

 CAUTION

Do not apply excessive force to the phase adjustment screw. Doing so may damage the 

internal variable capacitor.

IM 702907-01EN  2/2