5. Test Blocks

 

WARNING

•  If you set SW1 to SW4 to 500 mA when Vbus is supplying the LOAD TEST connectors 

(CN7 to CN9), the load resistors will become extremely hot. Make sure you do not touch 

the load resistors.

•  Since the load resistors of the LOAD TEST blocks get hot during the Droop and Drop 

tests, after finishing these tests, immediately remove the USB cables connected to 

connectors CN7–CN10 and disconnect the Vbus power supply from the upstream port, or 

turn OFF SW1–SW4.

LOAD TEST

  Four blocks (1 to 4) are available for the LOAD test. You can switch the load current between 100 

mA, OFF, and 500 mA using the switch for each of the blocks (SW1 to SW4).

Block Name

Connector Used

Switch Used

LOAD TEST[1]

CN7 (receptacle B)

SW1 (100 mA/OFF/500 mA)

LOAD TEST[2]

CN8 (receptacle B)

SW2 (100 mA/OFF/500 mA)

LOAD TEST[3]

CN9 (receptacle B)

SW3 (100 mA/OFF/500 mA)

LOAD TEST[4]

CN10 (receptacle B)

SW4 (100 mA/OFF/500 mA)

Note

Supply a voltage of 5 V ±5% to Vbus.

DROOP TEST

  Switches the load current between 0 mA and 100 mA at cycles of approximately 2 Hz.

  Connector used: CN5 (receptacle B)

INRUSH TEST

  SW5 switches between TEST ON, OFF, and DISCHARGE.

  When measuring the inrush current, attach a current probe to the current measurement loop.

  Cable and connector used: CN16 (cable A) and CN20 (receptacle A)

Note

Do not apply a current of 1 Arms or greater to Vbus.

TRIGGER

  Cable and connector used: CN11 (cable A) and CN15 (receptacle A)

HOST SQ TEST

  Cable used: CN2 (cable A)

DEVICE RECEIVER SENSITIVITY TEST

  SW8 switches RL1(RL2) between INIT and TEST.

•  When Set to INIT

  The D+ and D− signals are routed between CN30 and CN25.

•  When Set to TEST

  The D+ signal is routed between CN29 and CN25, and the D− signal is routed between CN28 

and CN25.

  Vbus is always connected between CN25 and CN30 regardless of the SW8 setting.

  Cable and connector used: CN25 (cable B, or mini B cable with the /MN1 option) and CN30 

(receptacle B)

DEVICE SQ TEST

  SW8 switches RL2 (RL1) between INIT and TEST.

•  When Set to INIT

  The D+ and D– signals are routed between CN34 and CN31.

•  When Set to TEST

  The D+ and D– signals are routed between CN31 and the 45 Ω terminator.

  Vbus is always connected between CN31 and CN34 regardless of the SW8 setting.

  Cable and connector used: CN31 (cable B, or mini B cable with the /MN1 option) and CN34 

(receptacle B)

BACKDRIVE VOLTAGE TEST

  Measure D+, D−, and Vbus voltage at CN18.

  Connector used: CN13 (receptacle A)

AVERAGE CURRENT DRAW TEST

  This test block is used for “Device Framework Testing.” Measure the Vbus current between TP3 and 

TP4.

  For testing methods, see “C. Device Framework Testing” in the Universal Serial Bus Implementers 

Forum Full and Low Speed Electrical And Interoperability Compliance Test Procedure.

6. Preparing for Waveform Measurement

Affixing the Probe Attachment

•  PBA2500 Active Probe (Model 701913, Sold Separately)

  Affix the attachment (model B8080JP or B8080JQ) that came with the probe as shown in the 

diagram below.

•  PBD2000 Active Probe (Model 701923, Sold Separately)

  Affix the attachments (model B8080JP, 2 pcs.) that came with the probe as shown in the diagram 

below.

 

 

B8080JP

B8080JQ

PBA2500 

Active Probe

Signal input terminal

GND terminal

 

+

-

B8080JP

PBD2000 

Differential Probe

 

CAUTION

When inserting a probe with the attachment affixed into the waveform measurement 

connector, do not apply vibration or shock to the probe.

Waveform Measurement Connector

•  As for the waveform measurement connectors (CN3, CN12, CN17, CN18, CN26, and CN32), 

insert the accessory connectors into the connectors already installed on the test fixture in a two-

layer configuration. The connectors come in two layers upon shipment.

 

 

Fixture

Waveform measurement connectors

Accessory connectors stacked 

in two layers

•  Affix the dedicated attachments to the PBA2500 and PBD2000 probes, then insert into the 

specified pins of the second layer of connectors. For the locations and numbers of pins, see the 

procedure manuals for the individual tests.

 

 

LED3

[ TEST

 ]

CN34

RL2

[ INIT ]

LED2

CN32

CN35

GND

[ DEVICE SQ

 TEST ]

CN31

CN30

CN29

RL1

CN28

CN2

6

E RECEIVE

R

IVITY TEST

1

1

CN2

5

G

D-

D+

G

G

D-

D+

G

D-

VBUS

LED3

[ TEST

 ]

CN34

RL2

[ INIT ]

LED2

CN32

CN35

GND

[ DEVICE SQ

 TEST ]

CN31

CN30

CN29

RL1

CN28

CN2

6

E RECEIVE

R

IVITY TEST

1

1

CN2

5

G

D-

D+

G

G

D-

D+

G

D-

VBUS

+ -

PBA2500 

Active Probe

PBD2000 

Differential Probe

 

CAUTION

Do not insert anything into the waveform measurement connectors other than the PBA2500 

or PBD2000 with dedicated attachments affixed, or the accessory sockets.

Note

Never pull out the top layer of connectors except when changing them.

7. Troubleshooting

Troubleshooting

  If servicing is necessary, or if the fixture is not operating correctly after performing the corrective 

actions below, contact your nearest YOKOGAWA dealer.

Description

Possible Problem

Corrective Action

LED1, 2, or 3 does not turn ON.  Using an AC adapter of a 

voltage outside the rated range.

Use the AC adapter that came 

with the fixture. 

Faulty connection between the 

AC adapter and the CN100 

connector of the test fixture.

Check the connection between 

the AC adapter and connector 

CN100 of the test fixture.

Trigger does not activate on the 

inrush current waveform.

SW5 not set to TEST ON.

Set SW5 to TEST ON when 

measuring. 

The VID, PID, connected 

address, and port of the DUT 

(host/hub) are not displayed in 

the HS Electrical Test Tool.

SW8 not set to INIT.

Set SW8 to INIT.

8. Specification

Safety Standards

Conforming standards EN61010-1

Overvoltage category (installation category) II*
Pollution degree 2**

*    Overvoltage category (installation category) is a number that defines excessive voltage, and 

includes regulations for impulse withstand voltage. II applies to electrical equipment that is 

powered by a fixed installation such as a distribution board.

**  Pollution degree refers to the degree of adherence by a solid, liquid, or vapor that reduces 

the withstand voltage or surface resistance factor. Pollution degree 2 applies to normal indoor 

atmospheres (with non-conductive pollution) only.

Emmisions

Conforming standards EN61326-1  Class A

EN61326-2-1
EN55011  Class A, Group 1
C-tick  EN55011  Class A, Group 1

This product is a Class A (for industrial environment) product. Operation of this product in a 

residential area may cause radio interference in which case the user is required to correct the 

interference.

Cable conditions
•  Use a 5 m USB cable at CN20 (receptacle A) on the INRUSH TEST block.
•  Use a 5 m USB cable at CN34 (receptacle B) on the DEVICE SQ TEST block.
•  Use a 1 m USB cable at the A and B receptacles on the other TEST blocks.

Immunity

Conforming standards EN61326-1 Table 2 (for use in industrial locations)

EN61326-2-1

Cable conditions
•  Use a 5 m USB cable at CN20 (receptacle A) on the INRUSH TEST block.
•  Use a 5 m USB cable at CN34 (receptacle B) on the DEVICE SQ TEST block.
•  Use a 1 m USB cable at the A and B receptacles on the other TEST blocks.

Reference operating 

conditions

Ambient temperature

23˚C ± 5˚C

Ambient humidity

55 ± 10% RH

Error in supply voltage 

and frequency

Within 1% of rating

Storage Environment

Temperature

–20˚C to 60˚C

Humidity

20% to 80% RH (no condensation)

Operating environment

Temperature

5˚C to 40˚C

Humidity

20% to 80% RH (no condensation)

Storage altitude

3000 m or less

Operating altitude

2000 m or less, indoors

Rated supply voltage

100 to 240 VAC

Permitted supply voltage range

90 to 264 VAC

Rated supply frequency

50/60 Hz

Permitted supply frequency range

45 to 66 Hz

Test fixture input voltage

5 V ± 5%

Test fixture power fuse

1.1 A time lag, UL/CSA compliant

Maximum power consumption

Approximately 7 VA

Withstand voltage (AC adapter, between input 

terminal and output terminal)

3.0 kVAC for one minute

Insulation resistance (AC adapter, between input 

terminal and output terminal)

500 VDC, 100 M

Ω

 or more

External dimensions

Approximately 317 (W) x 31 (H) x 75 (D) mm, 

excluding projections

Weight

Approximately 430 g, excluding AC adapter

Standard Accessories

AC adapter (A1626UP)

1

Power cord

1

Waveform Measurement Connector (B8080YH)

6

PBD2000 probe pin, B8080JP (variable type)

6

PBA2500 probe pin,  B8080JP (variable type) 

B8080JQ (straight type)

4
4

Carrying case (B8080JL)

1

User's manuals (this manual)

1

•  Only with the /F30 option

Software (CD-ROM)

1

Software user’s manual (IM701985-61E)

1

The procedures for device, host, and hub are included in PDF format on the software CD-ROM.

IM 701985-01E  3rd Edition  2/2