5. Test Blocks
WARNING
• If you set SW1 to SW4 to 500 mA when Vbus is supplying the LOAD TEST connectors
(CN7 to CN9), the load resistors will become extremely hot. Make sure you do not touch
the load resistors.
• Since the load resistors of the LOAD TEST blocks get hot during the Droop and Drop
tests, after finishing these tests, immediately remove the USB cables connected to
connectors CN7–CN10 and disconnect the Vbus power supply from the upstream port, or
turn OFF SW1–SW4.
LOAD TEST
Four blocks (1 to 4) are available for the LOAD test. You can switch the load current between 100
mA, OFF, and 500 mA using the switch for each of the blocks (SW1 to SW4).
Block Name
Connector Used
Switch Used
LOAD TEST[1]
CN7 (receptacle B)
SW1 (100 mA/OFF/500 mA)
LOAD TEST[2]
CN8 (receptacle B)
SW2 (100 mA/OFF/500 mA)
LOAD TEST[3]
CN9 (receptacle B)
SW3 (100 mA/OFF/500 mA)
LOAD TEST[4]
CN10 (receptacle B)
SW4 (100 mA/OFF/500 mA)
Note
Supply a voltage of 5 V ±5% to Vbus.
DROOP TEST
Switches the load current between 0 mA and 100 mA at cycles of approximately 2 Hz.
Connector used: CN5 (receptacle B)
INRUSH TEST
SW5 switches between TEST ON, OFF, and DISCHARGE.
When measuring the inrush current, attach a current probe to the current measurement loop.
Cable and connector used: CN16 (cable A) and CN20 (receptacle A)
Note
Do not apply a current of 1 Arms or greater to Vbus.
TRIGGER
Cable and connector used: CN11 (cable A) and CN15 (receptacle A)
HOST SQ TEST
Cable used: CN2 (cable A)
DEVICE RECEIVER SENSITIVITY TEST
SW8 switches RL1(RL2) between INIT and TEST.
• When Set to INIT
The D+ and D− signals are routed between CN30 and CN25.
• When Set to TEST
The D+ signal is routed between CN29 and CN25, and the D− signal is routed between CN28
and CN25.
Vbus is always connected between CN25 and CN30 regardless of the SW8 setting.
Cable and connector used: CN25 (cable B, or mini B cable with the /MN1 option) and CN30
(receptacle B)
DEVICE SQ TEST
SW8 switches RL2 (RL1) between INIT and TEST.
• When Set to INIT
The D+ and D– signals are routed between CN34 and CN31.
• When Set to TEST
The D+ and D– signals are routed between CN31 and the 45 Ω terminator.
Vbus is always connected between CN31 and CN34 regardless of the SW8 setting.
Cable and connector used: CN31 (cable B, or mini B cable with the /MN1 option) and CN34
(receptacle B)
BACKDRIVE VOLTAGE TEST
Measure D+, D−, and Vbus voltage at CN18.
Connector used: CN13 (receptacle A)
AVERAGE CURRENT DRAW TEST
This test block is used for “Device Framework Testing.” Measure the Vbus current between TP3 and
TP4.
For testing methods, see “C. Device Framework Testing” in the Universal Serial Bus Implementers
Forum Full and Low Speed Electrical And Interoperability Compliance Test Procedure.
6. Preparing for Waveform Measurement
Affixing the Probe Attachment
• PBA2500 Active Probe (Model 701913, Sold Separately)
Affix the attachment (model B8080JP or B8080JQ) that came with the probe as shown in the
diagram below.
• PBD2000 Active Probe (Model 701923, Sold Separately)
Affix the attachments (model B8080JP, 2 pcs.) that came with the probe as shown in the diagram
below.
B8080JP
B8080JQ
PBA2500
Active Probe
Signal input terminal
GND terminal
+
-
B8080JP
PBD2000
Differential Probe
CAUTION
When inserting a probe with the attachment affixed into the waveform measurement
connector, do not apply vibration or shock to the probe.
Waveform Measurement Connector
• As for the waveform measurement connectors (CN3, CN12, CN17, CN18, CN26, and CN32),
insert the accessory connectors into the connectors already installed on the test fixture in a two-
layer configuration. The connectors come in two layers upon shipment.
Fixture
Waveform measurement connectors
Accessory connectors stacked
in two layers
• Affix the dedicated attachments to the PBA2500 and PBD2000 probes, then insert into the
specified pins of the second layer of connectors. For the locations and numbers of pins, see the
procedure manuals for the individual tests.
LED3
[ TEST
]
CN34
RL2
[ INIT ]
LED2
CN32
CN35
GND
[ DEVICE SQ
TEST ]
CN31
CN30
CN29
RL1
CN28
CN2
6
E RECEIVE
R
IVITY TEST
1
1
CN2
5
G
D-
D+
G
G
D-
D+
G
D-
VBUS
LED3
[ TEST
]
CN34
RL2
[ INIT ]
LED2
CN32
CN35
GND
[ DEVICE SQ
TEST ]
CN31
CN30
CN29
RL1
CN28
CN2
6
E RECEIVE
R
IVITY TEST
1
1
CN2
5
G
D-
D+
G
G
D-
D+
G
D-
VBUS
+ -
PBA2500
Active Probe
PBD2000
Differential Probe
CAUTION
Do not insert anything into the waveform measurement connectors other than the PBA2500
or PBD2000 with dedicated attachments affixed, or the accessory sockets.
Note
Never pull out the top layer of connectors except when changing them.
7. Troubleshooting
Troubleshooting
If servicing is necessary, or if the fixture is not operating correctly after performing the corrective
actions below, contact your nearest YOKOGAWA dealer.
Description
Possible Problem
Corrective Action
LED1, 2, or 3 does not turn ON. Using an AC adapter of a
voltage outside the rated range.
Use the AC adapter that came
with the fixture.
Faulty connection between the
AC adapter and the CN100
connector of the test fixture.
Check the connection between
the AC adapter and connector
CN100 of the test fixture.
Trigger does not activate on the
inrush current waveform.
SW5 not set to TEST ON.
Set SW5 to TEST ON when
measuring.
The VID, PID, connected
address, and port of the DUT
(host/hub) are not displayed in
the HS Electrical Test Tool.
SW8 not set to INIT.
Set SW8 to INIT.
8. Specification
Safety Standards
Conforming standards EN61010-1
Overvoltage category (installation category) II*
Pollution degree 2**
* Overvoltage category (installation category) is a number that defines excessive voltage, and
includes regulations for impulse withstand voltage. II applies to electrical equipment that is
powered by a fixed installation such as a distribution board.
** Pollution degree refers to the degree of adherence by a solid, liquid, or vapor that reduces
the withstand voltage or surface resistance factor. Pollution degree 2 applies to normal indoor
atmospheres (with non-conductive pollution) only.
Emmisions
Conforming standards EN61326-1 Class A
EN61326-2-1
EN55011 Class A, Group 1
C-tick EN55011 Class A, Group 1
This product is a Class A (for industrial environment) product. Operation of this product in a
residential area may cause radio interference in which case the user is required to correct the
interference.
Cable conditions
• Use a 5 m USB cable at CN20 (receptacle A) on the INRUSH TEST block.
• Use a 5 m USB cable at CN34 (receptacle B) on the DEVICE SQ TEST block.
• Use a 1 m USB cable at the A and B receptacles on the other TEST blocks.
Immunity
Conforming standards EN61326-1 Table 2 (for use in industrial locations)
EN61326-2-1
Cable conditions
• Use a 5 m USB cable at CN20 (receptacle A) on the INRUSH TEST block.
• Use a 5 m USB cable at CN34 (receptacle B) on the DEVICE SQ TEST block.
• Use a 1 m USB cable at the A and B receptacles on the other TEST blocks.
Reference operating
conditions
Ambient temperature
23˚C ± 5˚C
Ambient humidity
55 ± 10% RH
Error in supply voltage
and frequency
Within 1% of rating
Storage Environment
Temperature
–20˚C to 60˚C
Humidity
20% to 80% RH (no condensation)
Operating environment
Temperature
5˚C to 40˚C
Humidity
20% to 80% RH (no condensation)
Storage altitude
3000 m or less
Operating altitude
2000 m or less, indoors
Rated supply voltage
100 to 240 VAC
Permitted supply voltage range
90 to 264 VAC
Rated supply frequency
50/60 Hz
Permitted supply frequency range
45 to 66 Hz
Test fixture input voltage
5 V ± 5%
Test fixture power fuse
1.1 A time lag, UL/CSA compliant
Maximum power consumption
Approximately 7 VA
Withstand voltage (AC adapter, between input
terminal and output terminal)
3.0 kVAC for one minute
Insulation resistance (AC adapter, between input
terminal and output terminal)
500 VDC, 100 M
Ω
or more
External dimensions
Approximately 317 (W) x 31 (H) x 75 (D) mm,
excluding projections
Weight
Approximately 430 g, excluding AC adapter
Standard Accessories
AC adapter (A1626UP)
1
Power cord
1
Waveform Measurement Connector (B8080YH)
6
PBD2000 probe pin, B8080JP (variable type)
6
PBA2500 probe pin, B8080JP (variable type)
B8080JQ (straight type)
4
4
Carrying case (B8080JL)
1
User's manuals (this manual)
1
• Only with the /F30 option
Software (CD-ROM)
1
Software user’s manual (IM701985-61E)
1
The procedures for device, host, and hub are included in PDF format on the software CD-ROM.
IM 701985-01E 3rd Edition 2/2