17
IM 701933-01E
Specifications
Product Specifications
Accuracy of 23 ±3 °C, 30 minutes after turning ON the power.
Frequency band*
DC to 50 MHz (–3 dB)
(see the typical characteristics shown on the next page, figure 1)
Rise time*
7 ns or less
Maximum continuous input
30 Arms (AC and DC components)
(See the next page, figure 2 for derating according to frequency)
Maximum peak current
50 Apeak, non-continuous
Output voltage rate*
0.1 V/A
Amplitude accuracy*
0 to 30 Arms: ±1.0% of reading ±1 mV
30 Arms to 50 Apeak: ±2.0% of reading
(DC, 45 to 66 Hz)
Noise*
Equivalent to 2.5 mArms or less
(for 20 MHz band measuring instrument)
Insertion impedance
(See typical characteristics on the next page, figure 3)
Temperature coefficient for sensitivity*
± 2% (with 50 Hz, 30 Arms input in a range of 0 to 40 °C)
Propagation delay
13 ns (Typical)
Maximum rated power
5.6 VA (within maximum input)
Rated supply voltage
±12 V ±0.5 V
Operating temperature and humidity
0 to 40 °C, 80% RH or less (no condensation)
Storage temperature and humidity
–10 to 50 °C, 80% RH or less (no condensation)
Operating location
Indoor use, altitude up to 2000 m
Effect of external magnetic fields
Equivalent to a maximum of 20 mA
(in a DC or 60 Hz, 400 A/m magnetic field)
Measurable conductors
Insulated conductors
Diameter of measurable conductors
φ 5 mm
Guaranteed accuracy period
1 year (10000 open and close operations)
Cable lengths
Sensor cable: approx. 1.5 m
Power supply cable: approx. 1 m
External dimensions
Sensor: approx. 175 (W) × 18 (H) × 40 (D) mm
Terminator: approx. 27 (W) × 55 (H) × 18 (D) mm
Weight
Approx. 230 g
Accessories
User’s manual, soft case
* In conjunction with a waveform measuring instrument having an input impedance of 1
MΩ ±1%