10-58
IM 701730-01E
Measurement Parameters
Twenty-six types of parameters and delay between waveforms can be selected as
operands of computation parameters. For details on the measurement parameters, see
section 10.6.
Computation Parameters
• Four parameters, User1 to User4, can be computed and displayed.
• Computation can be performed by setting the measurement parameters of the two
areas of the measurement target waveform as operands and using ope, -, *,
or /. The results are displayed as computation parameters User1 to User4.
• Equation
For example, the equation of computation parameter User1 is as follows:
User1 = M1 operator M2
M1:
Measurement parameter of Area 1 of the measurement target waveform
M2:
Measurement parameter of Area 2 of the measurement target waveform
Operator: +, -, *, or /
The measurement target waveforms of M1 and M2 can be selected from CH1 to
CH4(2)*, Math1, Math2, None or Const1 to Const4.
*
On the DL1735E/DL1740E/DL1740EL, you can select from CH1 to CH4, Math1, Math2, None
or Const1 to Const4. On the DL1720E, you can select from CH1 to CH2, Math1, None or
Const1 to Const2.
• If one of the measurement target waveform is set to None, the measured value of
the other measurement parameter is displayed.
• If both measurement target waveforms are set to None, the measurement value
display area shows ....
Turning the Computation Parameters ON and OFF
The display of the computation parameters User1 to User4 can be turned ON/OFF
separately.
ON:
Displays the measurement parameter.
OFF: Does not display the measurement parameter.
Unit of Computation Parameters
Units can be assigned to computation parameters User1 to User4 using up to four
characters.
• The type of characters that can be used are those displayed on the keyboard.
• The specified unit is displayed when the display of the measurement parameter is
turned ON.
10.8 Performing Automated Measurements of Waveform Parameters on Dual Areas