IM 2769-10E
―
8
―
(4) When the galvanometer indication comes to zero through adjustment of the measuring dial,
the resistance value of Rx is calculated from the following equation.
Rx =
Indication on
the measuring dial
Multiplying
factor
×
[
Ω
]
(5) When measurement is finished, first release the GA button switch.
Then, turn off the BA switch and set GA sensitivity dial to
GA
OFF
position.
(6) When it is desired to measure the resistance below 1 mΩ to 0.1 mΩ,
connect the 2771 standard resistor to the instrument as shown in Fig. 4.1.
In this case, the plug should be removed.
The unknown resistance Rx is calculated from the following equation.
Rx =
Dial
indication
Value of external
standard resistor
×
[
Ω
]
1
100
2769
Portable Double Bridge
2771
Standard Resistor
0.01
Ω
Rx
C
2
MULTIPLY
MAX CURRENT (A)
×10
0.01
0.1
0.3
1
3
×1
×0.1 ×0.01 ×0.001
EXT BA
C
2S
C
2
P
2
C
1
P
1
P
2S
P
1S
Rx
Fig. 4.1 Connection for External Standard Resistor