(c) When an external power source unit is used, be careful not to burn resistive elements by
over-current.
The allowable electric power of this instrument is 0.2 W (continuous) which dose not
degrade accuracy of the resistive element.
In case a power source unit is used of large capacity, connect a fuse 0.5 A in
series with the power source.
(d) The maximum allowable input of the internal galvanometer is 5 V.
When a high voltage external power source unit is used,
first roughly balance the galvanometer with a low voltage, and then increase the voltage.
It is essential to complete the measurement as soon as possible.
Be sure not set No. 1 dial of Rs to 0, and also not to short the Rx terminals.