Guidance and manufacturer´s declaration – electromagnetic immunity
The Model YM201/YM301 are intended for use in the electromagnetic environment specified
below. The customer or the user of the Model YM201/YM301 should assure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
± 8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power supply lines
100 kHz repetition frequency
± 1 kV for input/output lines
N/A
N/A
Surge
IEC 61000-4-5
± 0.5 kV,
± 1 kV
differential
mode line-line
N/A
N/A
Voltage
dips,short
interruptions and
voltage variations
on power supply
0 % UT (100 % dip in UT )
for 0.5 cycle at 0°, 45°, 90°,
135°,180°, 225°, 270°, and
315°
N/A
N/A