39
V2.3.0 OM-UNN valid from 15
th
April 2015
6.2
Technical Data
Index
Performance
Sample Positions
30
Modular Control
1 to 10 analyzers can be connected to a single computer.
Sample volume
Minimum 20μL (check instruction for use of each individual kit for the
proper amount of sample to be used)
Reaction Temperature
25
℃
to 37ºC±1
℃
Assay Method
EIA
Assay Speed
(60
~
120)min/30 tests
Calibration
Calibrator Included in all reagent kits
Quality Control
Internal QC material Included in all reagent kits
Test Combination
Different tests with the same reaction temperature can be performed
simultaneously in one cycle
Light Source
LED
,
3.6V / 0.504W
Filters
450nm / 630nm
OD Resolution
0.001ABS
Range of OD
0.000
~
3.000ABS
Accuracy of Filters
≤ ±2nm
Half Width of Filters
8 ±2nm
Linearity of OD
0.000~3.0000A, r≥0.98
Repeatability of OD
CV≤5%
Accuracy and
Fluctuation of Temperature
When operating at 25
℃
, offset ≤±0.5
℃,
Fluctuation≤±0.1
℃
Carry-over
≤1%
Accuracy and Precision of
Priming
Offset≤
5
%,
CV≤2
%
Noise
<80 dB
Dimension
680mm×600mm×500mm
Weight
48 kg