P2: Auto

-

Shut

-

Off Time

 

• 

From the parameter '

Set  1

' if the 

'SET'

 key is pressed once, the user 

gets access to parameter 

'Set 2'

. ’

SET  2

’ 

allows the user the ability to 

adjust  the  auto

-

shut

-

off  time.  The 

'UNIT'

  key  decreases  the  current 

value and the 

'FUNC' 

key increases it.

 

• 

The allowed values for this parameter are in the range from 1minute 

to 30 minutes in 1 minute increments.

 

• 

To disable this auto

-

off feature (turned on until a manual turn

-

off) the 

user must increase or decrease the parameter value until the  

'OFF'

 

value is displayed.

 

 

P3: Buzzer on Alarm Enable/Disable

 

• 

From the parameter '

Set  2

' if the 

'SET'

 key is pressed once more, the 

user gets access to parameter 

'Set 3'

 which allows the user the ability 

to enable or disable the buzzer sound in case of an alarm.

 

• 

Press the 

'UNIT' 

or the 

'FUNC'

 key to change the actual value.

 

• 

The  detector  is  set  from  the  factory  with  the  alarm  buzzer  enabled 
(Set 3=On).

 

• 

Pressing  the 

'SET'

  key  again  will  result  in  exiting  the  parameters 

configuration mode. When exiting the detector will  store all changes 

in the detector

s memory as the preferences.

 

 

OVER

-

RANGE FUNCTION

 

• 

This  function 

is 

continuously 

checking  the  measured 

concentrations  to  verify  that  they  remain  within  the  detector

s  full 

range limits. 

 

• 

When the full range value is exceeded, the internal buzzer will start 

beeping at the maximum repetition rate and the 

'ALM'

 symbol will 

be  turned  on  and  the  lower  display  will  show  the  label 

'OFL' 

(overflow)

.

 

 

LOW BATTERY FUNCTION

 

• 

The  detector  monitors  the  battery  level  and  when  it  is  below  a 
certain pre

-

set value, the low

-

level battery icon '

' is displayed 

in the upper right side of the screen.

 

• 

When this symbol is turned

-

on, the detector is still operational.

 

• 

When the battery voltage falls below a critical level, the words 

'Lo 

Batt'

  are  shown  on  the  display  and  the  detector  stops  operation.  

The only action allowed to the user is the turn

-

off.

 

• 

When the detector is turned on again and the battery voltage is in 
the valid range, all the functions are enabled.

 

• 

For  battery  replacement  please  follow  the  directions  explained  in 
the paragraph 'Beginning use'.

 

 

BUZZER

 

• 

This detector has an internal buzzer for the indication of  measured 

concentration and as a sound feedback on some user actions.

 

• 

When  the  detector  is  detecting  a  gas  concentration  greater  than 

zero,  the  rate  of  the  'beeps'  increases  according  to  the 

concentration level.

 

 

SENSOR FAIL

 

• 

The  detector  periodically  checks  the  sensor  status.  When  the 

sensor  is  found  to  be  defective,  the  buzzer  sound  is  set  to 

continuous  (when  the  Buzzer  Alarm  feature  is  Enabled)  and  the 

words 

'SEnS FAIL'

 are shown on the display.

 

 

TECHNICAL FEATURES

 

Power supply:

 

9V, 6x1.5V AAA alkaline batteries

 

External power supply:

 

12 Vcc, 100 mA

 

Battery life:

 

4 hours minimum

 

Battery level:

 

6.00 V (instr. operational)

 

Low

-

Batt level:

 

5.50 V (instr. stopped)

 

Sensor type:

 

Semiconductor

 

Pre

-

heating time:

 

max. 45 seconds fixed

 

Auto

-

shut

-

off time:

 

OFF, 1

-

 30 minutes

 

Display:

 

LCD TN

 

Protection grade:

 

IP 20

 

Operating temperature:

 

0°C .. +40°C

 

 

Storage temperature:

 

-

10°C .. +50°C

 

Humidity limits:

 

20% .. 80% RH

 

 

non condensating

 

Enclosure:

 

 

Material:

 

      

 

        ABS HB self

-

extinguishing 

 

 

Sensor Failure

 

When  the  detector  sensor  is  found  to  be  defective,  the 

buzzer  sound  is  on  continuously  (when  the  Buzzer  Alarm 

feature is  Enabled)  and  the  words 

'SEnS  FAIL'

  are  shown 

on the display.

 

1

 

Remove sensor sticker from protective cover at the end 
of the wand (see figure 4)

 

2

 

Unscrew sensor protective cover

 

3

 

Pull sensor straight out

 

4

 

Align tab in end of the wand with the tab on the sensor 
and insert sensor until fully seated

 

5

 

Screw on sensor protective cover

 

6

 

Apply new sensor sticker

 

7

 

Turn on unit and allow unit to heat and resume testing

 

 

Color:             Blue

 

Dimensions:   2.8x5.9x1.5in (72 x 151 x 37mm). (W x H x D) 

 

Weight:           ~.7lbs (~ 312 gr.)

 

Sensor Wand Flexible Length:  10.6in (270 mm)

 

Accessories:  External adaptor 230V~/12V= 100mA 

(available upon request)

 

 

WARRANTY

 

 

The 

Combustible Gas  Leak  Detector 

is guaranteed for 

12 

months

 from purchasing document date.

 

 

YELLOW  JACKET

® 

undertakes  to  repair  or  replace, free  of 

charge, those parts that, in its opinion, are found to be faulty 

during  the  warranty  period.  The  products  which  are  found 

defective  during  the  above mentioned periods of time must 

be  delivered  to  YELLOW  JACKET

®  

postage  paid.  The 

following cases are not covered by this warranty: accidental 

breakage  due  to  transport,  inappropriate  use  or  use  that 

does  not  comply  with  the  indications  in  the  product

instruction manual or quick start guide. 

 

Any  mistreatment,  repairs  and  modifications  to  the  product 

not  explicitly  authorized  by  YELLOW  JACKET

® 

shall 

invalidate the present warranty. 

 

 

Contact  YELLOW  JACKET

® 

Customer  Service  for  a  return 

request form and shipping address.

 

Fig. 4:

 

Sensor change

 

Remove 

sticker  

and  unscrew  cap 
to remove sensor

 

Ritchie Engineering Co./YELLOW JACKET

®

 