66
2. Press “NPA” to open binocular testing window.
3. Place the visual chart in the distance of 40cm, make sure the patient see clearly the row sighting
marks.
4. Press “+” and “-” to gradually adjust the spherical power until the sighting mark becomes
obscure. Press the obscure icon, and it is highlighted, then save the prismatic power of the obscure
point.
5. Press “+” and “-” to gradually adjust the spherical power until the sighting mark becomes distinct
again. Press the recover icon, and it is highlighted, then save the prismatic power of the recover
point.
5.6 Customer Self-programming
The device has standard optometry procedure A as factory default setting. There are four optometry
procedures can be programmed by the customer.
5.6.1 Self-programming
1. Press in the main interface to enter system parameter setup interface.
2. Select “Program” to enter self-programming interface.
Obscur
Recover