YATO YT-72305 Quick Start Manual Download Page 2

YT-72305

ELECTRONIC MIKROMETER

PL

1. measurement surfaces 

2. micrometer frame

3. LCD

4. mandrel knob               

5. friction clutch knob

6. control buttons     

7. fixing clamp

Technical data 

Measurement range:  

0 - 25 mm

Resolution: 

0.001 mm / 0.00005 inches 

Precision: 

±0.002 mm / ±0.001 inch 

Power of measurement:  

5 - 10 N

Display:  

LCD 

Electric energy consumption: 

 35 

μ

A

Power supply:  

one silver battery 1.55V, type SR44

Operating temperature:  

5

O

C - 40

O

C / 41

O

F - 104

O

F

Storage temperature:  

-20

O

C ÷ 60

O

C / -4

O

F ÷ 140

O

C   

Humidity:  

With no importance below 80% of relative humidity 

Absolute measurement 

Loosen the fixing clamp.

Turn the device on, using the „ON” button. Select the unit of measurement, pressing the „ABS/INC•••UNIT” button and holding it for approximately two seconds. Each time the result of the measurement 

will be alternately displayed in inches and millimetres. The result in inches is indicated with an „in” in the display. 

Press the „ABS/INC•••UNIT” button, until „ABS” is displayed.

Separate the measurement surfaces to a distance which slightly exceeds the width of the object to be measured. Then using the mandrel knob turn to adjust the distance roughly, and then continue 

tightening, turning the friction clutch knob. Stop tightening when the clutch has engaged. Read the result of the measurement from the display.

Relative measurement

The measurement principle is as the one mentioned above, but the purpose of relative measurements is to compare two measurements. Press the „ABS/INC•••UNIT” button to reset the display, which is 

indicated with an „INC” symbol. Once another measurement has been realised, a difference of dimensions may be read. 

The micrometer is equipped with a steel ball in a rubber fixture, which permits installation on the measurement surface of the micrometer. It facilitates measurements of small surface elements. Once it 

has been installed, it is required to reset the display in accordance with the aforementioned procedure. 

Attention! It is recommended to fix the object to be measured or the micrometer before the measurement has commenced. It will result in a higher precision of measurements.

It is recommended to realise a couple of measurements in order to obtain an arithmetic average of the results.

Maintenance 

The device should always be dry and clean (liquids may damage its moving parts).

The measurement surfaces should be cleaned with care, using a cotton cloth. Do not ever use gasoline, acetone or other solvents.

In order to save energy turn the device off, if it is not going to be operated for a prolonged period. The micrometer is equipped with an automatic deactivation system which is triggered after five minutes 

of inactivity. 

Do not ever apply any electric load to any element of the micrometer and do not use an electronic pen, which might damage the electronic circuits of the device.

Replacement of the battery 

The battery symbol in the display indicates low voltage of the battery. Open the battery compartment using a coin or the instrument provided, and then replace the battery, maintaining the correct polarity.

Troubleshooting

Defect

Cause    

Remedy

„E 1” in the display 

Excess of data 

Return the measurement surfaces to the original position or press the „ON/OFF···SET” button in 

order to restore the initial conditions

„E 3” in the display

Error or failure of the measurement sensor Remove the battery and install it anew; if the problem does not disappear, have the device repaired 

Incorrect  result of the measurement

Dirty measurement surfaces or incorrect 

measurement mode 

Clean the measurement surfaces, return to the zero position, reset the display and repeat 

measurement

The display is not illuminated or the indication is 

not clear

Incorrect contact of the battery 

Battery voltage below 1.4V

Remove the battery lid and correct its position to ensure its proper contact.

Replace the battery

Preparation for work

Clean the measurement surfaces and the guide using a cotton cloth. Do not apply solvents and acids.

Verify the zero position, bringing the measurement surfaces into contact, turning using only the friction clutch knob. Then press the „ON/OFF•••SET” and hold it for approximately two seconds to reset 

the display.  

Attention! Do not tighten using the mandrel knob, which may lead to distorted measurements.   

Verify the proper operation of all the buttons, switches and the display.

Basic measurement methods

ENVIRONMENTAL PROTECTION

Correct disposal of this product: This marking shown on the product and its literature indicates this kind of 

product mustn’t be disposed with household wastes at the end of its working life in order to prevent possible 

harm to the environment or human health. Therefore the customers is invited to supply to the correct disposal, 

differentiating this product from other types of refusals and recycle it in responsible way, in order to re - use this 

components. The customer therefore is invited to contact the local supplier office for the relative information to 

the differentiated collection and the recycling of this type of product.