Issue: 03.2013 (EN)
19
Insert the blue
calibration disc
and take a
measurement.
–
Press the measuring head together until a beep can be
heard.
–
Compare the value measured with the value (3) on the
rear of the calibration disc.
The deviation must be no greater than the maximum
deviation indicated on the disc (4).
Calib. value:
63,4
Note:
If the value measured does not lie within the specified
tolerance range, a reliable fertilizing recommendation is not
possible. In such cases, please contact YARA directly (see
the reverse of this manual).