4
Handling Precautions
CAUTION!
Sample installation method
The holder for attaching sample is prepared as optional accessory for each separator funnel
rhoto, centrifugal tube, and test tube rack. Select the desired holder, and attach it to the
shaking stand of this unit.
In the case of shaking the holder for separator funnel rhoto, centrifugal tube, and test tube, insert
the positioning pin of the holder into the 2 holes of the pipe hole, and secure it tightly.
Preparing timer
This unit incorporates a 60-min meter for timer. If it is necessary to operate this unit for a longer
time, set the timer to CONT. Continuous operation is available.
The timer accuracy of this unit is around +/-15%. It is not suitable for the shaking that is
required precise time setting.
Return after power failure
When power is supplied after a power failure, the device automatically starts operation again with
the same state as just before the power failure. It is danger that the device starts unattached
operation after a power failure. We recommend for you to turn off the switch of this unit if a
power failure occurs during operation.