13
Installation Method
Connecting method and assembling procedures of glass unit
3) Piping method
Piping between condenser and peripheral device
The rotary evaporator requires the vacuum equipment such as vacuum pump or aspirator, cooling water
circulation system to cool the condenser, and tap water. Use the vacuum hose with the inner diameter
of 6mm for vacuum route and heat insulation hose with the inner diameter of 9mm for cooling route.
Securely fix the connection between respective hoses and resin nipple using a clamp.
3-1. Piping between RE300 and glass set condenser
Piping method to glass set A condenser
Piping method to glass set B condenser
Piping method to glass set C condenser
IN
OUT
Vacuum pump
IN
OUT
Cooling water
circulation system
Exhaust
To drain when connected to
tap water
IN
OUT
Vacuum pump
IN
OUT
Cooling water
circulation system
Exhaust
To drain when connected to tap water
IN
OUT
Vacuum pump
Connect the attached drain hose
To drain
Exhaust
Put dry ice or ice.