24
4. How to Use
④
The water tap in the hood can be opened/closed by using the remote control handle on the
control panel.
⑤
The power outlet on the side panel provides electric power source of AC 115V 15A for
apparatus.
W a r n i n g
● The power capacity of power outlet indicates the total capacity of two plugs.
If the given capacity is violated, ignition may be caused.
● Take care with electric shocks from the power outlet.
An electric shock accident may take place if the power plug outlet is
touched by wet hand, or a conductive material is inserted into the outlet.
● Always use an appliance with grounding terminal in its power plug.
If not grounded, ground leakage or fire may be caused.
⑥
Suction prevention net
In the lower side of baffle plate, there is a suction prevention net installed to prevent foreign
objects such as pieces of paper or vinyl from being sucked. If an object is adhered to the net,
stop the test temporarily and remove the object. Then, resume the test.
If an object is adhered to the net, the blockage of baffle plate tunnel makes it impossible to
discharge air properly.
⑦
Gas tap (optional)
The gas tap installed within the hood can be remotely controlled with the remote handle on
the control panel, provided the knob of the gas tap has been opened manually. After using
this product, always close both the gas tap in the hood and the remote handle.
⑧
Door opening degree limiting stopper (option)
This stopper limits the opening height of windowpane. Usually it is used in locked condition.
If the stopper is required to be disabled, for example for a work preparation, the windowpane
can be opened over the maximum open degree, by unlocking the stopper. After the required
work is completed, always lower the windowpane, lock the stopper in position, and take the
key to the person in charge for storing.