![Yamato FP 100 Instruction Manual Download Page 22](http://html.mh-extra.com/html/yamato/fp-100/fp-100_instruction-manual_905272022.webp)
20
Operation Method
Auto Stop Operation
Select "Wait Function"
§
The screen changes to the window to select “Wait Function” after the time/hour is set.
§
Turning on the function temporarily stops the countdown of timer if the in-furnace
temperature is lower than set temperature by 3
℃
or more, or higher by 6
℃
or more.
§
The function is available only when the timer mode is set to “Time”. When it is set to
“Clock”, go to the next step.
§
Turn the jog dial to select “ON” or “OFF” and then press it to determine.
Wait
OFF
Auto-stop OP.
Start operation
§
The auto stop operation starts after the wait function is determined. Blinking auto stop lamp
lights on and the sub indicator shows the set temperature and remaining time before
operation stop.
Setup Temp 1000℃↑
Stop,after 30min
STANDBY
END
HEATER
FAN
FIXED TEMP
AUTO START
AUTO STOP
PROGRAM
§
If the wait function is set, the sub indicator indicates “Waiting” with blinking if the
in-furnace temperature is lower than set temperature by 3
℃
or more, or higher by 6
℃
or
more.
Setup Temp
100
℃↑
Waiting
30min
Stop operation
§
The device stops operation when the remaining time shows “0”. The end lamp blinks when
the operation is stopped. The sub indicator shows the operation end time.
OP. stop time
2000‑09‑17 13:00
STANDBY
END
HEATER
FAN
FIXED TEMP
AUTO START
AUTO STOP
PROGRAM
Cancel operation
§
Press the power key to cancel the operation.