How to obtain items of the proper weight (Inspection checkweighing function [#01=2])
(0) At the user parameter #03, you can select a timing at which the Evaluation LED light comes on.
At the user parameter #07, you can select a timing at which the indicated value flashes according to the
evaluation.
(1) As with the normal weight measurement, the panel shows 0 at the start and the instrument is ready to
weigh items.
(Start of measurement)
(2) As you put items on the instrument, the Down Arrow sign comes on while the items are underweight.
(Underweight)
(3) When you put items of the proper weight, the Evaluation LED light comes on.
(Proper weight)
(4) When you put items exceeding the proper weight range, the Up Arrow sign comes on. Remove the
excess from the instrument.
(Overweight)
#07 Flashing of indicated value per evaluation
0: Disable flashing
(Factory setting)
1: Flash upon underweight
2: Flash upon proper weight
3: Flash upon overweight
4: Flash upon underweight or overweight
5: Synchronize with #03
#03 LED lighting timing
0: Disable LED
1: LED ON upon underweight
2: LED ON upon proper weight
(Factory setting)
3: LED ON upon overweight
4: LED ON upon underweight and
flashes upon overweight