Yamatake Corporation
Operation
Model HGC303 - Heat Value Gas Chromatograph
3-55
A description of the data-setting panel is given next.
(1) PV1 - 11 configuration
~Note
*These configuration methods are same as those for ISO mode. Refer
to “ (1) PV1 - 11 configuration” on page 3-33.
(2) PV12 - 20 configuration
PV12 to 15, 19 and 20 can be configured to set which outputs are to be transmitted to
the host control system.
Figure 3-22 Output configuration panel (GPA)
Possible configurations for each PV are as follows:
Reference conditions
Configuration data for heat value calculation can be selected
or input manually.
Total of raw conc.
Usually, the total of raw concentration is within 95-
105mol% during process gas analysis. If these values are
required to be changed, input a user defined value for both
high and low limit.
Normalization method Normalization method can be selected.
Default is “Standard normalization”.
Multi-stream function
setting
use or Not use of multi-stream function can be selected.
This setting is only available for application of HGC (ver-
sion 3.0 or later) and HDM (version 2.30 or later) combina-
tion.
Download to HGC
All configured data are downloaded to the model HGC303
by clicking on this button. When the message appears,
download operation has been successfully completed.
Return Menu
Exit from Configuration mode. Return to Main Menu
Table 3-26 Possible configurations of PV12-20 (GPA)
PV12
Real Gross HV (dry) (BTU/CF) [default]
or Real Gross HV (sat) (BTU/CF)
or Ideal Gross HV (dry) (BTU/CF)
or Ideal Gross HV (sat) (BTU/CF)
or Gross HV (dry) (BTU/lbm)
Table 3-25 Description of configuration mode display (GPA)
Display
Description
Summary of Contents for HGC303
Page 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Page 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Page 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Page 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Page 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Page 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...