At any time while the software is capturing data to a file it is possible to stop the data capture
by selecting
Stop CSV Capture
from the file menu.
5
CONCENTRATION VIEW
The ability to turn the sensor signals into a known concentration of gas is of course the
requirement of most gas detection systems. The gas concentration view of the KEMET Gas
Sensing Evaluation Tool is available by selecting it from the drop down
Options
menu.
Figure 13
– Options Menu
Figure 14
– Gas Concentration View
Once opened the initial output of the concentration form will be nonsensical information as the
system has not been calibrated. In order to calibrate the system please determine the
coefficients according to the Beer Lambert law.
Once the system has been calibrated then it is possible to save the coefficients of the modified
Beer Lambert law by selecting file then
Save Coefficients
.