MODBUS OPERATION
28
Communication Setup
The Modbus communication interface is designed to
work with a Modbus Host device like PLC’s, etc. The
host must support RTU (Remote Terminal Unit) mode
to communicate with the H-3553T. Table 5-1 shows
the default settings required to communicate with
the H-3553T through the RS-485 Modbus interface
port. For more information regarding Modbus RTU
communication protocol and specifications see
www.modbus.org.
The H-3553T supports a Modbus client protocol interface. Modbus is an industry standard field bus for
interconnecting Programmable Logic Controllers (PLCs), intelligent sensors and other devices. The H-3553T
communicates Modbus via the RS-485 serial port connections; see Chapter 2 Table 2-1 for wiring connections.
This chapter will focus on Modbus setup and operation using the RS-232 menu interface or the SDI-12 interface.
Note
: Modbus communication is only available when the system is not busy performing other requested
tasks (i.e. purging, automode with the rate set to zero, SDI-12 commands, etc.)
Setting
Default Setting
Baud Rate
9600
Data Bits
8
Stop Bits
1
Parity
Even
Protocol
RTU
Table 5-1: H-3553T RS-485 Modbus Comm Settings
Function Codes
Modbus Function Codes are codes that are part of the Modbus command that specify what is being requested
via the Modbus protocol. The H-3555 supports two Modbus function codes. Table 5-2 shows the supported
function codes.
Function Code
Description
# of Registers
03
Read Holding Registers
41
16
Write Multiple Registers 41
Table 5-2: H-3553T Supported Modbus Function Codes
Holding Registers
The H-3553T has 41Holding Registers; these registers contain all the needed data for setup and measuring
the H-3553T. Table 5-3 is a list of the holding registers available.
Register Description
Start Address
# of 16-bit Registers
Data Type
*ID String
0/ “0000”
16
Char String
Modbus Address
16 / “0010”
1
Short Integer
Stage Units
17 / “0011”
1
Short Integer
Reserved
18 / “0012”
1
Short Integer
Baudrate
19 / “0013”
1
Short Integer
Parity
20 / “0014”
1
Short Integer
Bubble Rate
21 / “0015”
1
Short Integer
Purge Pressure
22 / “0016”
1
Short Integer
Purge Sustain
23 / “0017”
1
Short Integer
Purge
24 / “0018”
1
Short Integer
Table 5-3: H-3553T Holding Registers