Maintenance / cleaning
11
ba76171e02
04/2019
In the event of calcification, diluted hydrochloric acid (HCl) is recom-
mended (0.01 mol/l HCl, 0.036%, pH 2).
Preparing the cleaning
solution
The cleaning solution for the filter plate can be made from different
detergent concentrates by diluting with water. Prepare the cleaning
solution according to the following table. It can be mixed in the cleaning
container.
Pre-cleaning
Prior to each chemical cleaning, pre-clean the membrane with the spe-
cial brush and rinse it with water. See section 5.2 M
.
Chemical cleaning
1
Dismount the filter module from the guide slide.
2
Complete submerse the filter module (filter plate in the frame) in
the cleaning container with the cleaning solution for the filter
plate.
3
Start with a reaction time of 30 minutes.
Base solution
*
* Suppliers of the source solutions: see operating manual of your ana-
lyzer.
Preparation instruc-
tions
Yield
Household hygienic
cleaner based on
sodium hypochlorite
Add sufficient water to 1.5
l hygienic cleaner so that
the volume is 6 l
6 l
Technical sodium
hypochlorite solu-
tion (13 % active
chlorine)
Add sufficient water to
300 ml solution so that the
volume is 5 l
5 l
WARNING
Dangerous chemicals.
Improper use of chemicals can cause damage to your
health.
Heed the following rules:
Read the labels of the chemicals containers and follow
the safety instructions.
Wear your personal protective equipment (lab coat,
protective goggles, chemical resistant protective
gloves).