System 182-XT-4
Relay outputs
6 - 25
ba76027e02
12/2014
6.5
Behavior of linked relay outputs
6.5.1
Behavior in case of error
For linked relay outputs, you can specify the behavior of the outputs in
case of errors. Depending on the use of the output, the behavior in case
of errors is set in the following menus:
The specified behavior occurs with the following errors.
The system is in the maintenance condition or displays Init, ----, Cal,
Clean, Error or OFL.
The meaning of these displays is given in section 4.2 of this
operating manual.
The supply voltage is too low.
6.5.2
Behavior in non-operative condition
An output is non-operative when no function is activated for the output.
The relay contact is open in the non-operative condition.
An output becomes non-operative in case of
Power failure
(As soon as the supply voltage is sufficient again, the non-operative
condition of the outputs ends. The outputs function as specified by
the user again.)
Erasing a link to a sensor
Changing the Messmodus sensor setting for a linked sensor
Changing the Messbereich sensor setting for a linked sensor
Output
Menu
Limit indicator
Behavior at error(see section 6.4.4)
Frequency output
Frequency with error(see section 6.4.5)
Pulse-width output
Error pulse width(see section 6.4.6)
Before editing sensor settings a note appears on the
display to inform you that links will be erased when you
change the Messmodus or Messbereich sensor setting.
Summary of Contents for YSI 182-XT-4
Page 6: ...Contents System 182 XT 4 0 4 ba76027e02 12 2014...
Page 10: ...Overview System 182 XT 4 1 4 ba76027e02 12 2014...
Page 14: ...Safety System 182 XT 4 2 4 ba76027e02 12 2014...
Page 64: ...Installation System 182 XT 4 3 50 ba76027e02 12 2014...
Page 82: ...Operation System 182 XT 4 4 18 ba76027e02 12 2014...
Page 88: ...The SETTINGS menu System 182 XT 4 5 6 ba76027e02 12 2014...
Page 114: ...Relay outputs System 182 XT 4 6 26 ba76027e02 12 2014...
Page 116: ...Maintenance and cleaning System 182 XT 4 7 2 ba76027e02 12 2014...
Page 122: ...What to do if System 182 XT 4 8 6 ba76027e02 12 2014...
Page 134: ...Technical data System 182 XT 4 9 12 ba76027e02 12 2014...
Page 135: ...System 182 XT 4 1 ba76027e02 12 2014...
Page 138: ...Accessories and options System 182 XT 4 11 2 ba76027e02 12 2014...
Page 142: ...Appendix store separately if required System 182 XT 4 13 2 ba76027e02 12 2014...
Page 143: ......