General information
SenTix
®
HW-T 900(-P)
6
ba77090s02
01/2017
Carefully and using little power close the junction until the contact between
the rough ground surfaces can be noticed as a resistance. The electrode
is now ready to measure.
NOTE
Caution: Always use very little power to close the junction so the
ground surfaces are not damaged!
The ground junction can be opened further so that more electrolyte solu-
tion comes out.
Preparing the
sensor for
measurement
2.3
Calibration and measurement: General rules
Make sure the refilling opening for the reference electrolyte solution is
open.
Avoid the displacement of any solution (sample or buffer solution) from
one measurement to the next by taking the following measures:
– Shortly rinse the calibration and sample beakers with the solution the
beakers are to be filled with next.
– Between measurements, rinse the electrode with the solution that fol-
S
pecia
l w
renc
h
Ground
j
unction
O
pen
Cl
ose
SenTix
®
HW-T 900
Connect the sensor to the meter.
The sensor is immediately ready to measure.
SenTix
®
HW-T 900-P
Connect the sensor to a free IDS sensor plug-in
position of the multi parameter probe or to an IDS
connection of the meter. To open and close the
IDS plug-in position please note the section 2.1
O
.
The sensor is immediately ready to measure.
Connection cables in different lengths to connect
the SenTix
®
HW-T 900-P sensor to the meter are
.
Connect the sensor to the meter.
The sensor is immediately ready to measure.