General information
Automatic sensor
recognition
The sensor electronics with the stored sensor data are in the connecting
head of the electrode. The data include, among other things, the sensor type
and series number. In addition, the calibration data are stored in the sensor
with each calibration and the calibration history is recorded (the last 10
calibrations). The data is recalled by the meter when the sensor is connected
and is used for measurement and for measured value documentation.
Storing the calibration data in the sensor ensures that the correct slope and
asymmetry are automatically used if the sensor is operated with different
meters. On the other hand, different calibrated sensors can be used with one
meter without the need to recalibrate.
The digital transmission technique guarantees the failure-free
communication with the meter even with long connection cables. The sensor
firmware can be updated via the meter.