Xylem WTW SenTix H Operating Manual Download Page 6

Storage

SenTix

®

6

ba75739e02

07/2013

Subsequent cali-

brations

The frequency of subsequent calibrations depends on the application. Many 
meters provide an option where you can enter a calibration interval. After the 
calibration interval has expired, the meter will automatically remind you of the 
due calibration.

Storage

During short

measuring breaks

With the refilling opening open, immerse the combination electrode in 
reference electrolyte (KCl 3 mol/L, Ag

+

 free). Observe the minimum depth of 

immersion. Prior to the next measurement, shortly rinse the combination 
electrode with the test sample or deionized water.

Note

Prevent contact of the pH membrane to the beaker bottom to avoid scratches 
on the pH membrane.

Overnight or

longer

Put the clean combination electrode in the watering cap that is filled with 
reference electrolyte (KCl 3 mol/L, Ag

+

 free) and close the refilling opening. 

Note

pH combination electrodes must not be stored dry or in deionized water. The 
combination electrode could be permanently damaged by this. If the liquid in 
the watering cap has dried up, condition the combination electrode in 
reference electrolyte (KCl 3 mol/L, Ag+ 

+

 free) for at least 24 hours.

Note

During longer storing periods, salt sediments may develop on the watering 
cap. They do not affect the measuring characteristics and can easily be re-
moved with deionized water when the combination electrode is put into oper-
ation again.

Aging

Every pH combination electrode undergoes a natural aging process. With ag-
ing, the responding behavior becomes slower and slower and the electrode 
slope and asymmetry change. Moreover, extreme operating conditions can 
considerably shorten the lifetime of the combination electrode. These are:

Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins

High temperatures

High changes in pH and temperature.

The warranty does not cover failure caused by measuring conditions and 
mechanical damage.

Summary of Contents for WTW SenTix H

Page 1: ...OPERATING MANUAL ba75739e02 07 2013 SenTix PH COMBIN ELECTRODES WITH LIQUID REFERENCE SYSTEM AND GROUND JUNCTION SenTix H SenTix HW SenTix HWD...

Page 2: ...opyright Weilheim 2013 WTW GmbH Reproduction in whole or even in part is prohibited without the express written permission of WTW GmbH Weilheim Printed in Germany For the most recent version of the ma...

Page 3: ...round No Adjustable ground junc tion SenTix HWD KCl 3 mol L Ag free Ground Yes WTW model pH measur ing range Allowed tempera ture range Membrane resis tance at 25 C Typical application SenTix H 0 14 0...

Page 4: ...for the combination electrode to be stored Always keep the watering cap clean Connect the combination electrode to the meter Calibrate the combination electrode according to the operating manual of th...

Page 5: ...olyte solu tion comes out Calibration and measurement General rules Make sure the refilling opening for the reference electrolyte solution is open Avoid the displacement of any solution sample or buff...

Page 6: ...ning Note pH combination electrodes must not be stored dry or in deionized water The combination electrode could be permanently damaged by this If the liquid in the watering cap has dried up condition...

Page 7: ...the tip of the dropping bottle into the refilling opening while turning it slightly Pump the reference electrolyte in the shaft using the dropping bottle Pull the dropping bottle out of the refilling...

Page 8: ...lectrolyte solution for at least 1 hour Then recalibrate the combi nation electrode Wear parts and accessories Note Detailed information on our wide range of buffer solutions and more acces sories is...

Page 9: ......

Page 10: ......

Page 11: ......

Page 12: ...water needs Developing new technologies that will improve the way water is used conserved and re used in the future is central to our work We move treat analyze and return water to the environment and...

Reviews: