SenTix
®
950 / 980 / Micro 900(-P)
Technical data
9
ba75843e04
06/2017
After cleaning
Rinse the electrode with deionized water and condition it in reference elec-
trolyte solution for at least 1 hour. Then recalibrate the electrode.
6
Technical data
Measurement
pH measuring range
0.000 ... 14.000
Allowed temperature range
(°C)
0 ... 80 °C (SenTix
®
950(-P))
0 ... 100 °C (SenTix
®
980 / Micro 900(-P))
Allowed temperature range
(°F)
32 ... 176 °F (SenTix
®
950(-P))
32 ... 212 °F (SenTix
®
980 / Micro 900(-P))
Accuracy of the
IDS measuring
technique
Measured parameter
Accuracy (± 1 digit)
pH
± 0.004
U [mV]
± 0.2
T [°C]
± 0.1
General features
Reference electrolyte
KCl 3 mol/l, Ag
+
free
Junction
Ceramic
(SenTix
®
950(-P))
Platinum
(SenTix
®
980 / Micro 900(-P))
Temperature sensor
Integrated NTC 30
(30 k
at 25 °C / 77 °F)
Connection cable
Lengths
SenTix
®
950 / 980 / Micro 900: 1,5
SenTix
®
950 / 980 / Micro 900-P:
1,5 / 3 / 6 / 10 / 15 / 25 / 40 / 60 / 100 m
Diameter
4.3 mm
Smallest allowed
bend radius
Fixed installation:20 mm
Flexible use:60 mm
Plug type
Socket, 4 pins
Shaft dimensions,
material
Shaft length
120 mm
(SenTix
®
950 / 980(-P))
65/130 mm (SenTix
®
Micro 900(-P))
Shaft diameter
12 mm
(SenTix
®
950 / 980(-P))
12/5 mm
(SenTix
®
Micro 900(-P))
Shaft material
Polyamide (SenTix
®
950(-P))
Glass
(SenTix
®
980(-P),
SenTix
®
Micro 900(-P))