background image

Technical data

LS Flex/430

8

ba75968e02

11/2018

3

Technical data

Dimensions

approx. 236 x 82 x 170 mm

Weight

approx. 0.6 kg

Power

supply

Power pack 

FRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
-------------------

RiHuiDa RHD20W090150

-------------------

Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1.5 A
Connection max. overvoltage category II

Primary plugs contained in the scope of de-
livery: Euro, US, UK and Australian.

Serial

interface

Connection of the null modem cable (included in the scope of delivery)

Baud rate

Adjustable: 
1200, 2400, 4800, 9600, 19200 Baud

Type

RS232

Data bits

8

Stop bits

2

Parity

None

Handshake

RTS/CTS

Cable length 

Max. 15 m

Applicable directives

and standards

EMC

EC directive 89/336/EEC

EN 61326-1/A3:2003

FCC Class A

Meter safety

EEC directive 73/23/EEC

EN 61010-1 :2001

UL3101-1

CAN/CSA-C22.2 No. 1010.1-92

Climatic class

VDI/VDE 3540

IP protection class

EN 60529:1991

Summary of Contents for WTW LabStation LS Flex/430

Page 1: ...OPERATING MANUAL ba75968d02 11 2018 LS Flex 430 LabStation...

Page 2: ...ba75968d02 11 2018 LS Flex 430 Copyright 2018 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...LS Flex 430 Contents 3 ba75968e02 11 2018 Contents Overview 4 Battery operation with LS Flex 430 5 Technical data 8...

Page 4: ...o measurement is retained even after switching the pHotoFlex off and on again You can connect a bar code reader for the simplified calling up of programs The LSdata software included serves to easily...

Page 5: ...hile the meter is switched off Charging time of the rechargeable battery approx 36 hours CAUTION Use original power packs only The line voltage at the operating site must lie within the input volt age...

Page 6: ...screws 1 on the underside of the meter Remove the lid of the battery compartment 2 2 If necessary take four old batteries out of the battery compart ment 3 Connect the cable of the rechargeable batte...

Page 7: ...ar code reader see operating manual of your pHotoFlex The baud rate of the bar code reader is given in the operating manual of the bar code reader 4 Connect the power pack to an easily accessible powe...

Page 8: ...ined in the scope of de livery Euro US UK and Australian Serial interface Connection of the null modem cable included in the scope of delivery Baud rate Adjustable 1200 2400 4800 9600 19200 Baud Type...

Page 9: ......

Page 10: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: