Maintenance, cleaning, replacement
CellOx 325
8
ba25301e06
12/2017
5
Maintenance, cleaning, replacement
5.1
General maintenance notes
For your safety
Note the following safety instructions when handling electrolyte and cleaning
solutions:
CAUTION
The ELY/G electrolyte solution and RL-G cleaning solution irritate the
eyes and skin. Note the following points when dealing with these
solutions:
During working activities, always wear suitable protective gloves
and protective goggles/face shield.
If it comes into contact with the skin, rinse thoroughly with water and
immediately change contaminated clothing.
If it comes into contact with the eyes, rinse thoroughly with water
and consult a doctor.
Note the safety datasheets.
CAUTION
Before carrying out any maintenance activities, disconnect the sensor
from the instrument.
Note
Information on how to order wear parts and maintenance equipment can be
found in chapter 8 W
5.2
Outside cleaning
Cleaning agents
After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.
Contamination
Cleaning procedure
Lime sediments
Immerse in acetic acid for 1 minute (volume
share = 10 %)
Fat/oil
Clean with warm water that contains
washing-up liquid