Alyza IQ PO4
Lists
165
ba76201d03
05/2020
8
Lists
8.1
Explanation of the messages
This chapter contains a list of all message codes and the related message texts
that can occur in the log book of the IQ S
ENSOR
N
ET
system for the Alyza IQ an-
alyzer.
The last three digits of the message code form the component code. It identifies
the component (active component) that caused the message:
Some error messages contain an internal error code, starting with "#".
8.1.1 Error messages
Information on the contents and structure of the log book and the
structure of the message code is given in the L
OG
BOOK
chapter of
the IQ S
ENSOR
N
ET
system operating manual.
Module code
Component
3D1
Alyza IQ PO4, channel 1
3D2
Alyza IQ PO4, channel 2
552
MIQ/Alyza (adapter ADA)
Message code
Message text
EAM3Dx
Messbereich über- oder unterschritten
* Prozess überprüfen
EC1552
Fehler Automatische Kalibrierung: Kalibrierstandard konnte nicht
ermittelt werden, oder passt nicht zum eingestellten Messbereich.
Alyza IQ wurde gestoppt!
* check selected measuring range
* check calibration standard used
ES13Dx
Teilnehmer-Hardware defekt
* Service kontaktieren
ES2552
Teilnehmer-Hardware defekt xxx
* Service kontaktieren
ES3552
Gefahr durch Kondenswasser in der Messeinheit. Messeinheit
abgeschaltet.
* Messeinheitt entfeuchten
ES4552
Fehler Temperaturregelung der Messeinheit
* Service kontaktieren