31 

 

 

4  Connection of a relay box 

  General description 

4.1

The TW 7450 sample changer has a 9-pin plug connection for connecting pump and valve modules. Via this I/O 
connection, the devices available for this purpose can be (e.g. further MP 25 or a peristaltic pump) by means of 
an  accessory  cable.  The  I/O  connection  is  controlled  by  the  Titration  Software  TitriSoft.  There  are  4  outputs 
available.  The  necessary  commands  can  be  taken  from  the  command  list  (see  5  Command  list  of  the  sample 
changer). 

5  Command list of the sample changer 

The  TW  7450  sample  changer  has  a  number  of  commands  with  which  it  can  be  controlled  from  a  PC.  These 
commands  are  automatically  generated  by  TitriSoft.  If  you  want  to  create  your  own  application  programs,  the 
commands listed below must be used. 

The sample changer has no address and is connected directly to the PC via USB or the RS-232 cable supplied. 
The sample changer is controlled with the following RS parameters. 

Baudrate 

9600 

Parity 

No 

Start bits 

Stop bits 

 

Command 

Action 

abs = x-y-z 

The TW 7450 moves to the position with the coordinates x and y and lowers the head by 
the mm specified in parameter z. The x and y coordinates are specified in 0.1 mm steps, 
the z coordinate in 1 mm steps. The head moves up automatically before moving to a new 
position. 

getpos 

Returns the current position x-y-z in 0.01 mm units 

absz=z 

Lower the head by z mm in 0.01 mm units. The head can be lowered by max. 166 mm. 

up 

Lifting the head 

fc_sol=1 

Switch on the diaphragm pump connected to connection 7 (see Figure 6) 

fc_sol=0 

Switch off the diaphragm pump connected to connection 7 (see Figure 6) 

stiron 

Switching on the rod stirrer 

stiroff 

Switching off the rod stirrer 

stirsp=x 

Setting a stirring speed in x % in 1 % steps. The rod stirrer is not automatically switched on 
with this command. 

 

 

 

Summary of Contents for TW 7450

Page 1: ......

Page 2: ...Version dieser Gebrauchsanleitung finden Sie auf unserer Webseite Die deutsche Fassung ist die Originalversion und in allen technischen Daten bindend Operating Manual Page 19 34 Important notes The op...

Page 3: ...the TW 7450 25 2 2 System Preparations 26 2 3 2 3 1 Connection of the titration tower 26 2 3 2 Mounting the titration head 27 2 3 3 Electrodes titration tips stirrers 28 2 3 4 Using the sample changer...

Page 4: ...Copyright 2022 Xylem Analytics Germany GmbH Reprinting even as excerpts is only allowed with the explicit written authorization Germany Printed in Germany...

Page 5: ...The menu screens shown in this operating manual serve as an example and may differ from what you see Intended Use 1 2 The titration sample changer allows serial titrations with automatic sample change...

Page 6: ...er code TZ No 150 ml with spout low form 285216747 TZ 1784 150 ml with spout low form Super Duty 285202008 TZ 1788 250 ml with spout tall form 285201997 TZ 1787 250 ml with spout low Form Super Duty 2...

Page 7: ...g voltage is indicated on the specification plate underside of the device and backside of the power supply Nonobservance of this provision may result in damage to the titrator and the power supply or...

Page 8: ...e scope of delivery The unit can be placed on any flat surface Fig 1 Sample changer and cartons in delivery condition In the two boxes there are 4 plastic bags with various accessories Bag 1 with powe...

Page 9: ...e TW 7450 has the following connections 1 Power switch 2 Connecting the power supply unit 3 USB B connection for communication with the titration software TitriSoft 4 COM interface 2 not used 5 COM in...

Page 10: ...nger Fig 7 and fasten it with the four black screws Fig 8 Fig 7 and Fig 8 Mounting the titration tower Mount the cable guide of the titration tower on the rear side at the designated fixing points usi...

Page 11: ...wed together titration head Fig 12 is placed on the titration tower Fig 15 or Fig 16 by means of the pins and fixed by the screw 4 Option B for titration heads TZ 3969 and TZ 3974 2 3 2 2 The TZ 3974...

Page 12: ...stirrer socket on the back of the sample changer The stirring speed can be adjusted by software commands The cables required for the operation of the titration head are attached to the power supply o...

Page 13: ...the USB cable to a free USB A interface of the computer the USB driver is either automatically loaded via the Internet the PC must be connected to the Internet for this or the USB driver on the Titri...

Page 14: ...ameters are available on request Custom made racks for sample vessels with other diameters are available on request Fig 21 The sample racks are numbered with the corresponding positions and are loaded...

Page 15: ...ress and is connected directly to the PC via USB or the RS 232 cable supplied The sample changer is controlled with the following RS parameters Baudrate 9600 Parity No Start bits 8 Stop bits 1 Command...

Page 16: ...32 6 Coordinates of the sample rack Table 1 Coordinates of the 42 position rack Table 2 Coordinates of the 48 position rack...

Page 17: ...repaid 8 Storage and transportation If the titration sample changer TW 7450 has to be stored over some time or to be dislocated the use of the original packing will be the best protection of the devic...

Page 18: ...34 10 EC Declaration of Conformity The corresponding declaration of conformity of the device can be found on our homepage It will also be made available to you on request...

Page 19: ......

Page 20: ...ifi es pour le produit sont respect es Certificado del fabricante Certificamos que el aparato arriba mencionado ha sido controlado de acuerdo con la norma DIN EN ISO 9001 secci n 8 2 4 Seguimiento y m...

Reviews: