Replacing the dosing unit (installation and dissambly)

As a rule, the need for replacing the dosing unit occurs only rarely. The dosing unit has to be replaced, if
such a replacement becomes necessary as a result of a defect or of an inspection of the titration unit.

The dosing unit is equipped with lateral ribs around its circumference, with one of these ribs being in double
design. This double rib serves as a mark for the correct placement of the dosing unit (

&

 Fig. 3, Pos. 2 resp.

Fig. 9, Pos. 2).

Important note: The disassembly of the dosing unit as such is only possible using a special tool (TZ 3630)
and should only be done by experienced persons. The sealing lip of the PTFE piston and the sealings of the
liquid system may be damaged or incorrectly positioned in the course of such a disassembly. As a result, ti-
tration liquid would leak in unwanted places.

!

 

Caution! Wear protective glasses!

3.1 Piston Burette TITRONIC

â

â

â

â

 96

Ü

Switch the Piston Burette TITRONIC

â

 96  on (

&

 Fig. 8 of the 

Operating Instructions

  TITRONIC

â

 96)

Ü

Press the violet key of the manual key button, dose until the display shows between 18.00 and 19.00 ml.

Ü

Switch the Piston Burette TITRONIC

â

 96  off.

Ü

Unlock the dosing unit by rotating it to the left (anticlockwise, 

&

 Fig. 1).

Ü

Keep both keys of the manual key button depressed (

&

 Fig. 5 + Fig. 6), furthermore switch the titration 

unit on. (This somewhat awkward way of proceeding is necessary in order to prevent the inadvertent  start

of the process under all circumstances).

Ü

The motor of the Piston Burette TITRONIC

â

 96  will automatically move the piston rod up. The display  will

show <<uP>>.

Detaching the dosing unit:

Placement of the dosing unit:

Fig. 1  Unlocking

Fig. 2  Removal

Fig. 3  Placement

Fig. 4  Locking

Pos. 1

Pos. 2

Pos. 1

Pos. 2

Pos. 1

Pos. 2

M

Pos. 1

Pos. 2

Ü

The process is completed as soon as the display shows <<End>>, the dosing unit can be removed
(

&

 Fig. 2).

Ü

Switch the Piston Burette TITRONIC

â

 96  off.

Fig. 5

Fig. 6

Ü

Set on new (inspected, if necessary repaired) dosing unit vertically. When doing so the „M“ double rib
(

&

 Fig. 3) must point at the position 2 of the casing marking. Subsequently, switch the Piston Burette

TITRONIC

â

 96  on. The running motor will now pull the dosing unit downwards.

Ü

As soon as the dosing unit moving downwards has reached the edge of the casing, the Piston Burette
TITRONIC

â

 96  has to be switched off, and the dosing unit has to be locked manually (

&

 Fig. 4). In or-

der to fill the Piston Burette TITRONIC

â

 96  with titration solution, proceed as it is described in chapter

3.1 „Rinsing / Initial filling“ of the 

Operating Instructions

 (

&

 Fig. 9 and 10).

Summary of Contents for TITRONIC 96

Page 1: ...C 97 C O N T E N T S PAGE 1 Warning and safety information 10 2 Maintenance and care 11 3 Replacing the dosing unit installation and dissambly 12 3 1 Replacing the dosing unit of the Piston Burette TI...

Page 2: ...that a safe operation is no longer possible the packaging is damaged the titration unit shows visible damages the titration unit does not operate properly liquid has penetrated into the titration unit...

Page 3: ...o so remove the dosing unit chapter 3 Fig 1 to 4 resp Fig 7 to 10 If humidity can actually be found there one can conclude that the piston in the cylinder is no longer tight Check the hoses the thread...

Page 4: ...he dosing unit by rotating it to the left anticlockwise Fig 1 Keep both keys of the manual key button depressed Fig 5 Fig 6 furthermore switch the titration unit on This somewhat awkward way of procee...

Page 5: ...ther the time required for frequent changes is not more expensive than the acquisition of another dosing unit As a principle and in the case of all piston burette systems a substitution of the titrati...

Page 6: ...rty valve does not close Flush dosing unit replace solution When setting on a new dosing unit the piston rod is not properly pulled in the device rattles Inclined piston rod Set on piston rod vertical...

Page 7: ...6 pH electrode only for TitroLine 96 TZ 1509 Bottle set for titration agents TZ 2008 Bottle set on unit with S 40 thread for reagent bottles e g Merck make TZ 2005 Bottle set on unit with GL 45 thread...

Page 8: ...3455 Seal o ring 21 x 1 5 mm TZ 3500 Valve 20 ml Piston Burette TITRONIC 97 50 TZ 3150 Glass cylinder 50 ml TZ 3250 fitted with piston TZ 3450 TZ 3458 Seal o ring 33 x 1 5 mm TZ 3510 Valve 50 ml Orde...

Reviews:

Related manuals for TITRONIC 96