a) For each measurement, you must manually enter the measurement depth.
b) The mean station velocity (mean velocity in vertical) is calculated by integrating all
velocity measurements considering their location as illustrated in
Figure 5:5 - Vertical Velocity Curve
c) If multiple measurements are made at the same measurement depth, these
measurements are averaged prior to calculating the velocity integration.
Correction Factor
5.1.3.3
The
Correction Factor
(CF) is a user supplied parameter used to scale the mean
station velocity or velocity in the vertical.
Be careful when selecting a
Correction Factor
, as the mean station velocity is always
scaled by this value. Incorrect use of this parameter will affect the final discharge
measurement.
a) The
CF
is commonly used at edges, internal islands, and other method
None
stations to scale the velocity from the adjacent station(s).
b)
Correction Factors
implemented for each Station Type and
Velocity Method.
Table 5-3 – Correction Factors
Station Type
Velocity Method
Correction Factor
Bank
n/a
User defined
Island Edges
n/a
User defined
Open water
None
User defined
0.6
1
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...