![Xylem SonTek FlowTracker2 User Manual Download Page 136](http://html1.mh-extra.com/html/xylem/sontek-flowtracker2/sontek-flowtracker2_user-manual_3888944136.webp)
SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
135
g) Battery.
h)
To select a variable,
i).
Use the up or down arrow key to view a
variable.
i)
To accept the velocity measurement,
i).
Press the Right Soft Key.
j)
To
REDO
velocity measurement,
i).
Press the Left Soft Key.
The conventions and graphic display features used for displaying Velocity and SNR raw
data are defined under
Properties
associated with the variable output from the numerical calculations
performed on raw data collected is defined in
Table 8-8 - Review Point Measurement
Variable
Description
Units
Decimal
Samples
Total number of samples received.
number
n/a
Spikes
Number of spikes removed from mean velocity.
number
n/a
Vel
Mean X velocity component of all samples.
Velocity is despiked and mounting correction
applied (if applicable).
m/s or ft/s
4
σV
Standard error of X velocity component.
m/s or ft/s
4
SNR
Mean signal to noise ratio. Average of beam 1
and 2 SNR during point measurement.
dB
3
Angle
Flow angle relative to X direction. Angle
between the mean X velocity component and
mean Y velocity component.
Degree
2
Tilt
Mean wading rod angle during point velocity
measurement.
Degree
2
Temp
Mean temperature during point measurement.
°C or °F
2
Bnd
Boundary QC value.
n/a
n/a
Review Station Measurement
8.3.3.4
Review Station
measurement screen consist of tabular report and velocity distribution
plot supplied at the completion of all point velocity measurements at a station. The
variables supplied in the report are dependent on the features activated on the handheld,
velocity method selected and the discharge calculation method. The velocity distribution
plot gives graphical presentation of the point velocity measurements against theoretical
1/6
th
power law.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...