Quality Control
Criteria
QC Warning
Graphic Display
Peak Level
•
The shape of the sampling
volume curve is compared to
the expected shape. Any
significant deviation causes a
warning,
•
This criterion can only be
checked with sufficient SNR
(> 7 dB).
Peak Level >
QC
Peak Location
•
The physical location of the
sampling volume is compared
to the expected location. Any
significant deviation causes a
warning.
•
This criterion can only be
checked for sufficient SNR (>
7 dB).
Peak
Location >
QC
Temperature
•
No criteria defined for
Temperature.
•
Temperature data included in
Automated BeamCheck to
allow user to analyze data
before measurement.
None
c) The graphical display of the Quality Control parameters supplied during the
Automated BeamCheck is time series of each parameter. If there are any
significant variation between the beams, spikes are present in the data or Quality
Control warnings are issued we recommend the following steps,
i). Repeat the test at least once, after you verify that the probe and sampling
volume are well away from any underwater obstacles.
ii). Perform BeamCheck, described in
performance in more detail,
d) The display options for individual or all beam check data are listed in
The shortcut key on the keypad determines which beam SNR will be displayed,
Table 6-7 - Automated Beam Check Display
Beam
Line Color
Keypad Key
Beam1
Red
1
Beam2
Blue
2
Beam3
Green
3
All Beams
All
4
e)
To select a Quality Control Criteria,
i).
Use up or down scroll arrows keys to view quality control criteria.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...