You have the following options:
«
Init
»:
Starts a re-initialization of the TW 7650.
«
Get Vial
»:
Fetches a vial from the changer and places it in the oven.
The sample plate position is queried at the beginning.
«
Set Vial
»:
Take a vial out of the oven and place it in the changer.
The sample plate position is queried at the beginning.
«
Head Up
»:
Moves the needle head to the top position.
The gas supply must be switched separately
«
Head Down
»: Moves the needle head to the lowest position for sampling.
The gas supply must be switched separately.
Error messages
3.4
Error number
Description
Troubleshooting
401
The TO
7280 / TW
7650 is not ready for
operation
Check if the oven is switched on and
the cables are correctly connected to
the titrator and PC?
402
The TO
7280 / TW
7650 is in a non-initialized
state or stop state, e.g. after a work list has been
aborted.
The oven must be reinitialized manually.
404
Due to a device error, the system is in the stop
state.
The oven must be reinitialized manually.
If the error still occurs, please contact
our service department.
405
There is a serious internal device error
Please contact our service.