background image

Settings

HandyLab 600

46

ba77064e04

08/2021

7.3.3 Automatic 

Stability control

The automatic 

Stability control

 (AutoRead) function continuously checks the 

stability of the measurement signal. The stability has a considerable impact on 
the reproducibility of measured values. 

You can activate or switch off the automatic 

Stability control

 function (see 

section 7.3 S

ENSOR

-

INDEPENDENT

 

SETTINGS

, page 44).

The measured parameter flashes on the display

as soon as the measured value is outside the stability range 

when the automatic 

Stability control

 is switched off.

7.3.4 Automatic switch-off function

The instrument has an automatic switch-off function in order to save the 
batteries (see section 7.3.1 S

YSTEM

, page 44). The automatic switchoff function 

switches off the meter if no key is pressed for an adjustable period.

The automatic switchoff function is not active

if a USB-B cable is connected 

if the 

Automatic data storage

 function is active, or with 

automatic data trans-

mission

7.3.5 Display illumination

The meter automatically switches off the display illumination if no key is 
pressed for 20 seconds. 

The illumination is switched on with the next keystroke again.

You can also generally switch on the display illumination (see section 7.3.1 
S

YSTEM

, page 44).

Summary of Contents for SI Analytics HandyLab 600

Page 1: ...OPERATING MANUAL ba77064e04 08 2021 HandyLab 600 DIGITAL METER FOR DIGITAL IDS pH AND ORP SENSORS HandyLab 600...

Page 2: ...2 ba77064e04 08 2021 HandyLab 600 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany For the most recent version of the manual please visit www si analytics com...

Page 3: ...ents for safe operation 10 2 2 3 Unauthorized use 10 3 Commissioning 11 3 1 Scope of delivery 11 3 2 Power supply 11 3 3 Initial commissioning 11 3 3 1 Inserting the batteries 12 4 Operation 13 4 1 Ge...

Page 4: ...P calibration 39 7 Settings 40 7 1 pH measurement settings 40 7 1 1 Settings for pH measurements 40 7 1 2 Buffer sets for calibration 41 7 1 3 Calibration interval 43 7 2 ORP measurement settings 44 7...

Page 5: ...56 10 Maintenance cleaning disposal 57 10 1 Maintenance 57 10 1 1 General maintenance activities 57 10 1 2 Replacing the batteries 57 10 2 Cleaning 58 10 3 Packing 58 10 4 Disposal 58 11 What to do if...

Page 6: ...Contents HandyLab 600 6 ba77064e04 08 2021...

Page 7: ...andyLab 600 supports you in your work with the following functions Automatic sensor recognition Electronic access control Data transmission via the USB interface USB B 1 2 Sensors A measuring system r...

Page 8: ...ognition The automatic sensor recognition for IDS sensors allows to operate an IDS sensor with different meters without recalibrating to assign measurement data to an IDS sensor Measurement datasets a...

Page 9: ...might be damaged if the actions mentioned are not taken 2 1 2 Safety signs on the meter Note all labels information signs and safety symbols on the meter and in the battery compartment A warning symb...

Page 10: ...or safe operation The meter may only be operated according to the authorized use specified above The meter may only be supplied with power by the energy sources mentioned in this operating manual The...

Page 11: ...re MultiLab Importer 3 2 Power supply The HandyLab 600 is supplied with power in the following ways Battery operation 4 batteries 1 5 V Mignon type AA USB operation via a connected USB B cable 3 3 Ini...

Page 12: ...he signs on the batteries must correspond to the signs in the battery compartment You can operate the meter either with normal batteries or with rechargeable batteries Ni MH In order to charge the bat...

Page 13: ...tches the meter on or off M Selects the measured parameter Quits the settings CAL CAL_ Calls up the calibration procedure Displays the calibration data STO STO_ Saves a measured value manually Opens t...

Page 14: ...tion 5 Sensor symbol calibration evaluation calibration interval 6 Measured temperature with unit 7 Status information meter 8 Softkeys and date time AR Stability control AutoRead is active HOLD Measu...

Page 15: ...Info softkey 1 Digital sensors pH and ORP 2 USB B device interface 3 Service interface CAUTION Only connect sensors to the meter that cannot return any volt ages or currents that are not allowed SELV...

Page 16: ...eter performs a self test 2 Connect the sensor The meter is ready to measure 1 Switch the printer off with On Off 25 0 7 007 pH C 0 14 More 01 02 2014 08 00 BlueLine 24 pH IDS B092500013 pH resolution...

Page 17: ...selection is dis played with a frame Submenus The name of the submenu is displayed at the upper edge of the frame Sub menus are opened by confirming with ENTER Example Operating mode Explanation Meas...

Page 18: ...Calibration record function Messages Information is marked by the i symbol It cannot be selected Example Language Deutsch Audio signal off Illumination on Contrast 50 Shutoff time 1 h Temperature unit...

Page 19: ...g ENTER_ open the Storage config menu The instrument is in the setting mode 3 Select the System submenu with V W The current selection is displayed with a frame 4 Open the System submenu with ENTER 25...

Page 20: ...tion date Identification of stored measured values 5 Select the General submenu with V W The current selection is displayed with a frame 6 Open the General submenu with ENTER 7 Open the setting mode f...

Page 21: ...th V W and ENTER The setting menu for the date and time opens up 3 Select and confirm the Time menu with V W and ENTER The hours are highlighted 4 Change and confirm the setting with V W and ENTER The...

Page 22: ...e stability range when the automatic Stability control is switched off You can start the Stability control manually at any time irrespective of the set ting for automatic Stability control see section...

Page 23: ...tatus indicator appears while the measured value is assessed as not stable A progress bar is displayed and the display of the measured parameter flashes The HOLD AR status indicator appears as soon as...

Page 24: ...see section 7 1 1 SETTINGS FOR PH MEASUREMENTS page 40 Use one to five buffer solutions of the selected buffer set in any order When other buffer sets are used other nominal buffer values are display...

Page 25: ...r flashes 8 Wait for the end of the measurement with stability control or accept the calibration value with ENTER The calibration display for the next buffer appears voltage display 9 If necessary fin...

Page 26: ...ary finish the calibration procedure as a two point calibration with M The calibration record is displayed pH 0 0 24 8 C 7 000 01 07 2011 08 00 AR Buffer mV 15 Thoroughly rinse the sensor with deioniz...

Page 27: ...the calibration record is dis played The calibration line is determined by linear regression If single point calibration was set in the menu the calibration proce dure is automatically finished with...

Page 28: ...ltage display 10 If necessary finish the calibration procedure as a single point calibration with M The calibration record is displayed For single point calibration the instrument uses the Nernst slop...

Page 29: ...ssary finish the calibration procedure as a two point calibration with M The calibration record is displayed pH 4 035 24 8 C ConCal 2 01 07 2011 08 00 Buffer V W 18 Thoroughly rinse the sensor with de...

Page 30: ...calibration record is displayed or Continue calibrating using the next buffer with ENTER After the fifth buffer has been measured the calibration is automat ically finished Then the calibration record...

Page 31: ...arately The worse evaluation of both is taken into account The evaluation appears on the display and in the calibration record Menu item Setting function Explanation Calibration Calibration data stor...

Page 32: ...O DO IF page 59 For pH IDS sensors you can optionally enable a more finely graded calibration evaluation QSC see section 5 2 8 QSC FUNCTION SENSOR QUALITY CONTROL page 34 Display Calibration record Ze...

Page 33: ...e needle 2 Marking lines for all nominal buffer values used with the last valid cali bration 3 Measuring range for which a valid calibration is available Measured values in this range are suitable for...

Page 34: ...fer solutions It covers the measuring range of the sensor pH 2 to pH 11 The QSC initial calibration determines the actual condi tion of the sensor and stores it as a reference in the sensor To meet th...

Page 35: ...nsor evaluation with the sensor symbol for this sensor 1 Open the menu for measurement settings with ENTER 2 In the QSC menu select First calibration with V W The calibration display appears AutoCal Q...

Page 36: ...pH C 0 14 01 02 2014 08 00 Info USB output 1 Open the menu for measurement settings with ENTER 2 In the QSC menu select Control calibration with V W The calibration display appears AutoCal QSC Kit is...

Page 37: ...oon as the measured value is outside the stability range when the automatic Stability control is switched off You can start the Stability control manually at any time irrespective of the set ting for...

Page 38: ...sured value is assessed as not stable A progress bar is displayed and the display of the measured parameter flashes The HOLD AR status indicator appears as soon as a stable mea sured value is recogniz...

Page 39: ...39 6 2 ORP calibration yes 0 1 C Automatic with temperature sensor 1 C Manual ORP electrodes are not calibrated You can however check ORP electrodes by measuring the ORP of a test solution and compari...

Page 40: ...on More buffers and details see section 7 1 2 BUFFER SETS FOR CALIBRATION page 41 and section 5 2 PH CALIBRATION page 24 Calibration Single point calibration yes no Quick calibration with 1 buffer Cal...

Page 41: ...Resolution of the mV display Reset Resets all sensor settings to the delivery condition see sec tion 7 4 1 RESETTING THE MEASUREMENT SETTINGS page 47 Menu item Possible setting Setting Explanation No...

Page 42: ...08 25 C 12 Fisher 2 007 4 002 7 004 10 002 25 C 13 Fluka BS 4 006 6 984 8 957 25 C 14 Radiometer 1 678 4 005 7 000 9 180 25 C 15 Baker 4 006 6 991 10 008 25 C 16 Metrohm 3 996 7 003 8 999 25 C 17 Beck...

Page 43: ...mbol or the QSC scale flashes It is still possible to measure Setting the calibra tion interval The calibration interval is set to 7 days in the factory 21 Reagecon 20 2 000 4 000 7 000 10 000 13 000...

Page 44: ...ay After completing the settings switch to the measured value dis play with M Default settings are printed in bold 1 Open the menu for measurement settings with ENTER 2 In the Calibration Calibration...

Page 45: ...rate of the USB Device interface System Interface Out put format ASCII CSV Output format for data transmission For details see section 9 TRANSMITTING DATA USB INTERFACE page 55 System Interface Decima...

Page 46: ...ility control is switched off 7 3 4 Automatic switch off function The instrument has an automatic switch off function in order to save the batteries see section 7 3 1 SYSTEM page 44 The automatic swit...

Page 47: ...following settings for ORP measurements are reset to the default settings with the Reset function The sensor settings are reset under the Reset menu item in the menu for cali bration and measurement s...

Page 48: ...system settings are reset in the menu Storage config System Reset To open the Storage config menu press the ENTER_ key in the measured value display Setting Default settings Language English Audio sig...

Page 49: ...time output to the interface If the memory is full When all memory locations are occupied it is not possible to continue storing Then you can e g store the data from the memory to a PC see section 8...

Page 50: ...aphical display of the memory usage ID number 1 Interval 30 s Duration 180 min continue 0d03h00min 0 1d17h33min Automatic data storage 1 2 3 01 02 2014 08 00 Back Menu item Possible setting Explanatio...

Page 51: ...ows to switch off the automatic data storage function before the adjusted storage duration has expired 1 Remaining storage duration 2 Graphical display of the storage duration If the automatic data st...

Page 52: ...L or RCL_ key Settings The settings are explained here using the manual data memory as an example The same settings and functions are available for the automatic data memory Menu item Setting function...

Page 53: ...memory Erasing the measurement data memory see section 8 3 1 EDITING THE MEASUREMENT DATA MEMORY page 52 Manual data storage 3 of 64 01 02 2014 07 43 33 ID number 1 BlueLine 24 pH IDS B092500013 pH 7...

Page 54: ...AR appears with the measured value if the AutoRead crite rion was met while storing stable measured value Otherwise the AR dis play is missing Calibration evaluation 4 levels or no evaluation 8 3 4 St...

Page 55: ...tallation of the USB driver PC with at least one free USB connection and CD ROM drive Windows 2000 Windows XP Windows Vista or Windows 7 1 Output the current measurement data to the USB B interface wi...

Page 56: ...set the transmission interval Simultaneously with every automatic data storage process see section 8 2 AUTOMATIC DATA STORAGE AT INTER VALS page 50 Stored mea sured values Manual Displayed dataset wi...

Page 57: ...ce 1 Screws 2 Battery compartment 1 Unscrew the two screws 1 on the underside of the meter 2 Open the battery compartment 1 on the underside of the meter 3 Remove the batteries from the battery compar...

Page 58: ...ontact your supplier 4 Place four batteries type Mignon AA in the battery compartment 5 Close the battery compartment 2 and tighten the screws 1 6 Set the date and time see section 4 4 5 EXAMPLE 2 ON...

Page 59: ...r Use suitable IDS pH sensor Air bubble in front of the junction Remove air bubble e g sway or stir the solution Gel electrolyte dried out Replace the sensor Cause Remedy IDS pH sensor The values dete...

Page 60: ...cessary IDS pH sensor test sample Conductivity too low Use suitable IDS pH sensor Temperature too high Use suitable IDS pH sensor Organic liquids Use suitable IDS pH sensor Cause Remedy IDS pH sensor...

Page 61: ...undefined or EMC load unallowed Processor reset Press the ENTER and On Off key simultaneously Cause Remedy E g a question by the service department Switch on the meter Open the menu ENTER_ Stor age co...

Page 62: ...nditions Storage 25 C 65 C Operation 10 C 55 C Admissible relative humidity Yearly mean 75 30 days year 95 Other days 85 Power supply Batteries 4 x 1 5 V alkali manganese batteries type AA Operational...

Page 63: ...HandyLab 600 Technical data ba77064e04 08 2021 63 Meter safety EC directive 2006 95 EC EN 61010 1 IP protection class EN 60529...

Page 64: ...E VERSION IS IN THE METER OR IDS SENSOR PAGE 61 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the meter...

Page 65: ...R IDS SENSOR PAGE 61 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder already exists for the sensor or sensor type the new dat...

Page 66: ...activity divided by the unit of the molality The practical pH value is the value of a pH measurement Potentiometry Name of a measuring technique The signal depending on the mea sured parameter of the...

Page 67: ...on of the numerical value and unit e g 3 m 0 5 s 5 2 A 373 15 K Molality Molality is the quantity in Mol of a dissolved substance in 1000 g sol vent Reset Restoring the original condition of all setti...

Page 68: ...Glossary HandyLab 600 68 ba77064e04 08 2021...

Page 69: ...13 M Measured value display 17 Measurement accuracy 43 Measurement data memory Edit 52 Erase 52 Storage locations 54 Measurement dataset 54 Measuring ORP 37 pH 22 Menu for calibration and measurement...

Page 70: ...Index HandyLab 600 70 ba77064e04 08 2021...

Page 71: ......

Page 72: ...purpose creating advanced technology solutions to the world s water challenges Developing new technologies that will improve the way water is used conserved and re used in the future is central to ou...

Reviews: