Operation
DIQ/S 281
ba76226e02
03/2023
74
4.8
General course when calibrating, cleaning, servicing or
repairing an IQ sensor
When an IQ sensor is calibrated, cleaned, serviced or repaired, the mainte-
nance condition for the IQ sensor should always be switched on.
In the maintenance condition
the system does not react to the current measured value or the condition of
the selected IQ sensor
linked outputs are frozen
IQ sensor errors do not prompt changes in the conditions of linked outputs.
The maintenance condition for IQ sensors is automatically activated
during calibration. After calibration the IQ sensor remains in the maintenance
condition until the maintenance condition is switched off manually (see
section 4.8.3)
during a compressed air cleaning cycle.
Therefore, keep to the following course when you calibrate, clean, service or
figure 4-11
List of all components
with sensor info
6
Exit the
Status of output channels
window with
<M>
or
<ESC>
.