Tab 1 

Allowed calibration weights [g] for each 
model 

BL 303 

100 – 200 - 300 

BL 602 

200 – 400 - 600 

BL 2002 

1000 – 2000  

BL 3001 

1000 – 2000 - 3000 

BL 6001 

2000 – 4000 - 6000 

BL 30K1 

10000 – 20000 - 30000 

Examples: 

If the display shows the blinking value:  

2000, 

 put 

on the plate a certificated sample mass of 2000g. 

If the display shows the blinking value:  

100, 

 put 

on the plate a certificated sample mass of 100g.

 

Unit setting 

Unit selection

Press  

 button to select the desired unit. 

Default unit settings 

Default unit selected is gram  [g]. 

To change the default unit,turn on the balance, and 

keep pressed 

 button for 3 seconds. 

Display shows on the right the unit selected, and 
 

On

 if the unit is enabled, 

Off

 if It is disabled. 

to scroll units.

to switch 

On/

 

Off

Setting 

Off, 

that Unit will not be selectable in the measurment mode. 

To choose a new default unit, scroll the units and when is reached the desired one, press 

button .

 

Zero 

With no weight on the scales, press the 

button to take the 0.00 reading on the display. The ZERO function is only 

effective if the value shown is less than 4% of total capacity. If the value is greater than 4%, the load is regarded as a tare.  

Tare 

After placing the tare on the scales, press the

 button to reset the value displayed to 0.00. The load used for Tara has 

to be subtracted by the overall capacity weighing. 

Measure 

Place the sample on the plate. 
Wait for the stable indicator        on the left of the display before taking the measurement. 

N° 

Unit 

Symbol 

Conversion 

grammo 

Carat 

ct 

0,1999694 g 

Avorirdupois Pound 

lb 

453,59237 g 

Avorirdupois Ounce 

oz 

28,349523 g 

Troy Ounce 

ozt 

31,103477 g 

Avoirdupois Dram 

dr 

1,7718451 g 

Grain 

GN 

0,0647989 g 

Penny weight 

dwt 

1,5551738 g 

Momme 

mom 

3,749996   g 

10 

Hong K

ONG 

J

EWELRY TAEL

 

K.tl 

37,429002 g 

11 

Taiwan Tael 

T.tl 

37,49995   g 

12 

Hong Kong tael 

H.tl 

37,799375 g 

13 

Singapore tael 

S.tl 

37,799366 g 

14 

Tola 

11,663804 g 

15 

Mesghai 

4,6083162 g 

16 

Tical 

cl 

16,3293 g 

17 

Kilogram 

Kg 

1000 g 

18 

British Gravitational System 

B/G 

19 

PCS 

20 

21 

 

Summary of Contents for BL 303

Page 1: ...User Manual BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 Models...

Page 2: ...START 7 DATE AND TIME SETTINGS 7 USER CALIBRATION 7 CALIBRATION PROCEDURE 7 UNIT SETTING 8 UNIT SELECTION 8 DEFAULT UNIT SETTINGS 8 ZERO 8 TARE 8 MEASURE 8 COUNTER FUNCTION 9 RS 232 SETTINGS 9 ITEM SE...

Page 3: ...3 Glass protection BL 303...

Page 4: ...4 BL 602 1 2 3 4 5 6 7 8...

Page 5: ...5 BL 2002 BL 3001 BL 6001...

Page 6: ...Working Humidity 40 85 Installation and removal of the scale must be done with the machine switched off Do not use the scale upside down After switch On wait at least 30 minutes warming up before the...

Page 7: ...e value scroll to the next value confirm Successively are required 1 Month Month 2 Day Day 3 Hour Hour 4 Minute Min User Calibration Each balance is calibrated in the laboratory by technicians using c...

Page 8: ...ith no weight on the scales press the button to take the 0 00 reading on the display The ZERO function is only effective if the value shown is less than 4 of total capacity If the value is greater tha...

Page 9: ...he display shows first rS Set 8888888 According with the following table set the correct values a 0 9600 1 19200 b 0 print in continuos 1 print when stable 2 press to print c 0 1 print the date d 0 1...

Page 10: ...trument in correspondence of the load cell Datasheets Models BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 Cod 27000513 27000323 27000313 27000113 27000123 27000133 Capacity 300 g 600 g 2000 g 3000 g...

Page 11: ...of in public landfills According to the UE Directive 2002 96 EC the European users of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new one The il...

Reviews: