XO FLEX – Installation instructions
Installation
requirements
6
2
INSTALLATION REQUIREMENTS
2.1
SUPPLIES AND CONNECTIONS
Before installing the unit, please make sure that the installation supply requirements listed in Table
1 are fulfilled.
Table 1 – Installation requirements
Electrical
Requirement
Length above
floor surface
Mains supply
230 VAC + 10%, 50 Hz. PVC cable rated for >75° with earth and
min. 3 x 1.5 mm
2
. Main fuse: the electrical installation must be
secured with a 10 A fuse.
75 cm
Equipotential earth
(if required by
national law)
1 x 4.0 mm
2
Assistant call control
cable
Min. 2 x 0.1 mm
2
and max. 2 x 1.25 mm
2
Max. 2 A / 60 VDC or 2 A / 25 VAC
75 cm
X-ray unit attached
to the XO unit
Cable for X-Ray must have its own installations pipe. Cable shall
be connected to an installation box in the floor.
Suction motor
control cable
Min. 2 x 0.1 mm
2
and max. 2 x 1.25 mm
2
Max. 2 A / 60 VDC or 2 A / 25 VAC
75 cm
Positioning of cables
in the floor
See installation drawing YB-235.
Data
Requirement
Length above
floor surface
Isolation transformer When connecting the external PC to the unit, the external PC
must be equipped with a medical grade power supply or be
powered via a medical grade isolating transformer! XO part
number: XO-620.
RS-232 cable
Connect unit and external PC through a RS-232 cable,
male/female. Connect the cable’s male connector to the unit end.
70 cm
XO HD Display
Connect XO HD Display and an external PC through a HDMI
cable.
Depending on the environment, it might be necessary to use an
HDMI amplifier with the HDMI cable.
10 cm
Intraoral video
camera
Connect the intraoral video camera to an external PC through a
USB A cable male/female. The cable’s female connector is in the
unit end.
It is recommended to use a USB Extension 10 m cable with
Repeater.
USB
extension
cable: 20 cm
Suction, air and
water
Requirement
Height above
floor surface
Suction
Suction machine power >600 l/min.
Vacuum pressure at the connection point under static conditions:
Min = 35 mbar, Max = 150 mbar.
Plastic pipe Ø 32 mm with socket – see YB-235.
6 cm max.
Incoming
(compressed) air
Pipe 3/8” female internal thread – preferably fitted with a ball
valve – see YB-235.
7 cm max.