Xingyun XY-C Instruction Book Download Page 2

 

XYprecision electronic scale is a kind of intellectual measuring instrument with 

 

weighing, counting and other many functions. The product introduces imported Integrated 

circuit and enjoys stable and reliable performance, easy operation and accurate weighing 

and counting. As a kind of measuring instrument, it has been widely used by enterprises 

and scientific research institutions.

 

 Model specification and index 

 

Model

 

Specification

 

Readability

 

size of weighing tray

 

XY-C

 

 

110g-3100g

 

 

0.01g

 

 

 

Ф130mm

 

 

XY-B

 

 

110g-5100g

 

 

0.1g

 

 

 

Ф130mm

 

160*1

 

 Working Condition

 

Working Temperature:   0°

40°C

      

Maximum Consumed Power: 5W 

 

Temperature Fluctuation:   5°C/h

       

Power Supply: 220V, 60HZ

 

Comparative Humidity:     50

﹪~

85

 

 Operation Requirements

 

When it is operated, electronic scale shall be placed on stable working platform to avoid 

the influences from mechanical vibration, direct sunshine and air current etc.

 

 Operation

 

(1)

 

Power-on

 

(2)

 

Switch  on  the  power  supply  behind  the  scale  and  the  display  indicates  in  turn 

“8.8.8.8.8.8.” “Max weighing capability of scale”, “=, =, =, =, =”. At last it shall display 

the weighing mode as “XXX.XX”. If any error founded in display, you may turn on and 

turn off the scale once more. Then it will be operated normally.

 

 Calibration

 

(1)

 

Calibration Requirements

 

When distinct errors appeared in the weighing of scale, the scale shall be recalibrated to 

make  weighing  accurate.  The  recalibrated  scale  shall  be  placed  on  the  stable  working 

platform without the influences from the vibration of air current and strong electromagnetic 

wave.  The  calibration  results  will  be  more  accurate  20  minutes  later  after  the  scale  is 

operated. 

 

(2)

 

Calibration Procedure

 

One-point Calibration

 

Turn on the scale it shall display the weighing mode as “0.00”, then press the “Tare” button 

until shows “CAL”, then release. Display blinkingly the stand weight value and then place 

same value standard weight; The “= = = = = =” will be displayed; then display stably weight 

value  and  remove  weight.    then  “=  =  =  =  =  =”  will  be  displayed.  the  “0.00”  will  be 

displayed;. The calibration is finished.

 

Three-point Calibration

 

Turn off the scale, then press the “COU” button and turn on the scale simultaneously until 

the display shows “CAL”, then release. The calibration procedure shall be undertaken as 

blow:

 

Display  blinkingly  the  first  point  stand  weight  value  and  then  place  the  same  value 

standard weight;

 

Display stably weight value and then remove weight. The “= = = = = =” will be displayed;

 

Display  blinkingly  the  second  point  stand  weight  value  and  then  place  the  same  value 

standard weight;

 

Display stably weight value and then remove weight. The “= = = = = =” will be displayed;

 

Display  blinkingly  the  third  point  stand  weight  value  and  then  place  the  same  value 

standard weight;

 

Display stably weight value and then remove weight. The “0.00” will be displayed;

 

The calibration is finished.

 

 Weighing 

 

(1)

 

Turn on the scale to make it warm up stably and display “0.00”.

 

(2)

 

Place a thing on the weighing tray and display the weight of the thing. It indicates the 

weight is already stable when the display value is not changed again.

 

 Tare

 

(1)

 

If there is a container on the weighing tray, the scale displays the weight of container.

 

(2)

 

Press “TARE” and then “0.00” will be displayed. It indicates that the tare has been 

deducted. 

 

(3)

 

Place the thing into container and the scale will display the weight of thing. 

 

 The counting operation of scale 

 

(1)

 

Press “COU” button and “COU” will be displayed. Then display blinkingly one of value 

Reviews: