XIELI XL5145 Series User Introduction Manual Download Page 2

same 5V power, it will cause the series disturb.
4.Avoid strong electromagnetism interference, the measuring wire must use
shielding wire or twisted pair.
5.Avoid violent shake and impact.
6.Do regular checking, If you want to decrease its basic error while
performing checks, please remove back the cover, then use a small crew
driver slowly move the potentiometer through a hole which near by the
digital tube to make the reading match the standard value.

Simple Trouble Shooting

A. If the digital tube doesn’t light, please check weather the voltage of

power is normal.

B. If the digit is unstable, please check:

І

. Whether there is a strong electromagnetism interfering nearby.

. Whether the plug is well connected.

C. If the digit displays 0000, continuous twinkled, indicates the input signal

is beyond the full range, you should decrease its input voltage, and
short-circuit the input end, let the current go.

D. If the input end add the measured voltage, but the digit has no response,

please check weather the input connection is open circuit.

National Technical Supervisory Bureau Manufacture Calibration

Instrument License (Shaanxi) No.01000188

Shaanxi Xieli Photo Electric Instrument Co., LTD

Address

No. 2 Xinke Rd. Eastern Development Zone, Xi'an, Shaanxi, China

TEL

+86-29-82622375 FAX

+86-29-82623951 Postal code:710043

Email: [email protected]

Website

WWW.XIELI-CHINA.COM

XIELI

XL 5145 series

DIGITAL

PANEL

METERS

INTRODUCTION MANUAL

®

Reviews: