WALKAIDE II USER MANUAL
16
MNL-0023
Gait Mode
1) Start: The app sends the Gait mode parameters and the mode start command
to the stimulator, receives the stimulator
’s reply, and enters the walking
working interface.
2) Pause: The app sends a Gait mode pause (stop) command to the stimulator,
receives the stimulator
’s reply and pauses the walking.
3) Continue: The app sends the Gait mode (start), and you have to adjust the
intensity again. After receiving the stimulator
’s reply, the walking work
continues.
4) End: After long press and stop for 1.5 seconds, the Gait mode stop command
is sent to the stimulator; the app receives the stimulator
’s reply, and the Gait
mode stops.
Under the Gait mode for Pro Version, user can choose among smart mode, normal
mode and manual mode.
1) Smart Mode: The stimulator automatically calculates the tilt angle A to start
the electrical stimulation and the tilt angle B to end the electrical stimulation
according to the gait data of the first four steps of the patient. The parameters
hat can be adjusted are the electrical stimulation intensity, frequency and
pulse width in the lower-level interface of the
“Parameter Settings.”
2) Normal Mode: The stimulator performs electrical stimulation according to the
set parameters. The parameters that can be adjusted are the electrical
stimulation intensity, and the frequency, pulse width, tilt angle A, tilt angle B,
duration, delay time, ramp up and ramp down in the lower-level interface of
the
“Parameter Settings.”
3) Manual Mode: The clinician can manually press the
“Start” button to deliver
stimulation at certain time by observing the patient
’s gait when the patient is
walking. The parameters that can be adjusted are the electrical stimulation
intens-ity, frequency and pulse width in the lower-level interface of the
“Parameter Settings.”