2
2-14
Color Calibration
14.
Press and hold the measure button and wait for a beep.
N
OTE
:
To hear a beep with a Windows compatible computer, you must install a sound
card and speaker(s). You might need to turn up the volume on your computer to hear
the beep. Alternatively, you can configure the computer to display a visual signal for a
beep. Follow the procedure on
.
15.
After you hear a beep, slide the ES-1000 at a slow but consistent pace across the
strip.
Take about 5 seconds to slide the length of the strip.
16.
Release the button when all the patches in the strip have been measured, and you
have reached the white space at the end of the strip.
When a strip is measured successfully, the cross hair moves to the next color. If the strip
is not measured successfully, a message directs you to try again.
17.
Repeat
through
until all the strips have been measured.
N
OTE
:
You must take the measurements in the order outlined in the Status field.
18.
When all the patches have been read successfully, click Accept.
19.
Click Apply in the Calibrator window to implement the new calibration measurement.
20.
Click OK in the Information dialog box.
This completes the EX7750 calibration process.
Calibrating the EX7750 with the DTP41
Using the ColorWise Pro Tools Calibrator and the DTP41 spectrophotometer,
you can quickly measure color patches generated by the printer and automatically
download these measurements to the EX7750.
N
OTE
:
Before using ColorWise Pro Tools and the DTP41 to calibrate the EX7750,
follow the instructions on
for calibrating the DTP41 spectrophotometer.
Summary of Contents for Phaser EX7750
Page 1: ...OLOR UIDE...
Page 8: ...This page is intentionally blank...
Page 12: ...This page is intentionally blank...
Page 16: ...This page is intentionally blank...
Page 40: ...This page is intentionally blank...
Page 64: ...This page is intentionally blank...
Page 116: ...This page is intentionally blank...
Page 120: ...This page is intentionally blank...