![Xentaur XDT User Manual Download Page 32](http://html1.mh-extra.com/html/xentaur/xdt/xdt_user-manual_3273042032.webp)
Section 3: Instrument
25
•
The instrument will display the measurement using the newly acquired calibration values.
The instrument computes the dewpoint from the measured capacitance of the sensor. The
sequence of computations is as follows:
•
The capacitance is converted to percent of full scale (-100°C to +20°C dp), using the low dB,
low dew, and saturated dB when performing SpanCheck values.
•
Cal Adj is applied to correct the mid point of the curve.
•
If enabled, the multi-point calibration table is applied to further correct the transfer function.
•
If enabled, the single point calibration is applied to fine tune the transfer function.
•
The corrected percent of full scale is converted to dewpoint.
3.4.4.4 Viewing Serial Number Mode
In this mode the user can view the instrument serial number. The display shows the serial number
by alternately displaying
XSN
and the number. If the number is larger than 1999 then it is dis-
played in 2 parts, first part is the thousands signified by the x10
3
legend in the upper right corner
of the display and the second part is the units. For example serial number 12345 will be shown as:
Pressing the UP and DOWN buttons simultaneously resets the instrument, this is useful in instal-
lations where power can not be turned on and off easily. Pressing the Pressure Correct button will
cause the instrument to enter the setup state.
Pressing the MODE button, returns the unit to the Viewing Dewpoint Mode.
3.4.5 SetUp State
To enter the Setup State power-up the unit (or reset it from the serial number Mode), and hold the
MODE key pressed while the unit is initializing. Refer to Appendix B for a flow diagram of the
SetUp State.
The setup state provides eleven capabilities, each one controlled from its own mode. Depressing
the MODE button navigates through the modes in the following order:
n
Select Alternate Display
Units
o
Select Sensor Type
p
Set Measured Attenuation of Sensor at a Low Dewpoint
q
Set the
Low Dewpoint at which the Attenuation was measured
r
Set the Calibration Mid-range Adjust
s
Analog Output select 4/20mA or 0/24mA
t
Perform a Test and Verification of the Analog Out-
put
u
Set dewpoint corresponding to low end of Analog Output
v
Set dewpoint corresponding to
high end of Analog Output
w
Lock/Unlock the instrument
nn
Sensor Cable Compensation
(back to)
n
.
These eleven functions are explained in their respective sections which follow. Note that modes
6,7,8&9 will appear only on instruments with the analog output option. Changes can be made
only to an unlocked instrument, the changes are stored (when the MODE button is pressed) and
retained even if the instrument power is turned off.
1. Display of alternate units: In this mode, a second engineering unit can be chosen to be dis-
played alternating with the engineering unit selected in the operating state, for example, a
dewpoint can be alternately shown in °C and in PPM.
x10
3
Summary of Contents for XDT
Page 1: ...yyyyyyyyy yyyyyyyyy yyyyyyyyy yyyyyyyyy yyyyyyyyy yyyyyyyyy...
Page 9: ...XDT User s Manual 2...
Page 39: ...XDT User s Manual 32...
Page 48: ...XDT User s Manual Appendices 41 Appendix D Optional Sensor Fittings...
Page 51: ...XDT User s Manual Appendices 44 Appendix E continued XDT NEMA Enclosure Mounting Dimensions...
Page 65: ......